3
Setup and Installation
3-18
Narkomed GS Setup and Installation Manual
Draeger Medical, Inc.
Installing the
Fresh Gas
Oxygen
Sensor
Adapter
(Optional)
1. Hold the locking bar on the anesthesia machine's fresh gas outlet in
the extended position.
2. With the oxygen sensor port facing up, slide the notched portion of the
adapter stem into the retaining slot of the locking bar, so that the
adapter's 15 mm male fitting faces the anesthesia machine's fresh gas
outlet.
3. Release the locking bar, guiding the adapter's 15 mm male fitting into
the fresh gas outlet of the anesthesia machine.
4. Insert the oxygen sensor into the port on top of the adapter.
WARNING:
If the oxygen sensor is removed, the oxygen sensor port must
be closed with the oxygen sensor port plug.
5. Attach the non-rebreathing circuit fresh gas hose to the adapter's
barbed fitting. (For non-rebreathing circuits equipped with a 15 mm
fresh gas connector, cut off the connector and press the hose onto the
barbed fitting of the adapter.)
OXYGEN SENSOR PORT
FRESH GAS
LOCKING BAR
(EXTENDED POSITION)
FRESH GAS HOSE
BARBED FITTING FOR
NON-REBREATHING
GAS CIRCUIT
OXYGEN
SENSOR
FRESH GAS
OUTLET
OXYGEN
SENSOR PLUG
SU20997
ADAPTER
STEM
15MM MALE
FITTING
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...