Jaundice Meter (Model JM-103) User Manual (usr070)
Page 7 - 1
Section 7
Troubleshooting
Service Calls
When you call technical support about your unit, be prepared to give the
serial number from the product identification label. When you give the
serial number, the technical support representative can identify your unit
and give you the information you need more quickly.
Error Messages
For warnings that may appear on the display window, refer to the table
below.
Error Messages
Warning
Cause
Solution
Er1
The measured value is
abnormal. In the case of an
averaging measurement, the
measurement fluctuation is
excessively large.
Place the measuring probe
perpendicular to the infant’s
forehead (at hospital sites
only) or sternum (at hospital
sites or physicians’ offices),
and take the measurement
again. If
Er1
still appears,
contact the nearest authorized
service facility.
Er2
through
Er6
A measurement error may
have occurred during an aver-
aging measurement, or the
hardware is not functioning
properly.
Set the
Power
switch to the
Off
position, and then return
it to the
On
position. If the
warning still appears, contact
the nearest authorized service
facility.
Summary of Contents for JM-103
Page 2: ......
Page 24: ...DRAFT 20 June 2005 Page 2 16 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Page 30: ...DRAFT 18 May 2005 Page 3 6 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Page 44: ...DRAFT 20 June 2005 Page 5 8 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Page 48: ...DRAFT 18 May 2005 Page 6 4 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Page 68: ...DRAFT 20 June 2005 Page A 18 Jaundice Meter JM 103 User Manual usr070 NOTES...
Page 90: ...DRAFT 9 Nov 2004 Page C 14 Jaundice Meter JM 103 User Manual usr070 NOTES...