a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Dr trust Portable
ECG Smart Connect
- 1201
is used exceeds the applicable RF compliance level above, the
Dr trust Portable ECG Smart Connect
- 1201
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the Dr trust Portable ECG Smart Connect
- 1201
.
Recommended separation distances between portable and mobile RF communications
equipment and the Dr Trust Portable ECG Smart Connect- 1201.
Dr trust Portable ECG Smart Connect
- 1201
is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of
the Dr trust Portable ECG Smart Connect
- 1201
can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Dr trust Portable ECG Smart Connect
-
1201
as recommended below, according to the maximum output power of the
communications equipment.