When  the  SL-003  blower  is  used  in  an  electronic  assembly  area,  the 

ionized air stream should cover as much of the working area as possible.   

The  constant  flow  of  ionized  air  will  prevent  items  such  as  work 

surfaces,  tools,  materials  and  components  from  developing  a  static 

charge.  Charged  items  introduced  into  a  work  area  will  be  neutralized 

and will remain neutral while keeping in the ionized air stream. 

 

The  SL-003  overhead  blower  has  been  designed  for  minimal 

maintenance.  The  only  regular  maintenance  suggested  consists  of 

emitter  point  cleaning,  ion  balance  checking  and  ion  output  checking. 

Emitter point cleaning takes only moments with the point cleaners. The 

SL-003 also contains a balancing circuit that is inherently self balancing. 

This circuit compensates for emitter point wear, line voltage fluctuations 

and variations in air velocity. 
 
Scheduled checking of ion output and balance should be considered to 

assure quality audit requirements

5. MAINTENANCE  

5.1 Emitter Cleaning 

To clean ion emitter points, simply rotate point cleaner knob located at 

center of each outlet clockwise to the stop (approximately one turn) and 

release. The spring-loaded point cleaning brush will return to its home 

position. Recommended frequency of cleaning is once a week. 

 

5.2 Air Inlet and Outlet Cleaning 

The air inlet grilles on the top of the unit and ionized air outlets should 

remain clean to prevent restriction of airflow. They can be cleaned with 

a soft brush or vacuum. 

 

5.3 Ion Output Check 

Summary of Contents for SL-003

Page 1: ...1 INSTRUCTION SL 003 IONIZING AIR BLOWER DR SCHNEIDER PC...

Page 2: ...an overhead unit designed to provide total coverage for electrostatic problems Please read the instruction manual completely before installing or operating the device This device is equipped with a 3...

Page 3: ...olyurethane Foam Ion Balance offset voltage 0V 10V It requires no calibration 3 INSTALLATION The SL 003 Ionizing blower should be located 600 to 900mm above work surface with center fan directly above...

Page 4: ...performed by qualified personnel The color cord for wiring the plug is as follows Red line Blue Neutral Green Yellow Ground 4 OPERATION Activate the SL 003 Overhead Ionizing Air Blower by turning FAN...

Page 5: ...the point cleaners The SL 003 also contains a balancing circuit that is inherently self balancing This circuit compensates for emitter point wear line voltage fluctuations and variations in air veloci...

Page 6: ...ures are subject to the variation in the atmospheric conditions 4 Discharge times are in seconds from 1000 volts to 100 volts If a charged plate monitor is not available a hand held electrostatic fiel...

Page 7: ...not insert objects through intake or outlet grille Do not try to verify operation of unit by drawing a spark from an ion emitter point The design of the balancing circuit makes the park test inconclu...

Page 8: ...r flickers the compact fluorescent lamp should be replaced as soon as possible Access to lamp is provided by removing the front panel of the device Disconnect the device by unplugging line cord Remove...

Reviews: