background image

PosiTector  

DPM

®

Quick Guide v. 3.1 

Dew Point Meter

Summary of Contents for PosiTector

Page 1: ...PosiTector DPM Quick Guide v 3 1 Dew Point Meter...

Page 2: ...sured Ts Surface Temperature measured Td Dew Point calculated Ts Td Surface minus Dew Point Temperature calculated Tw Wet Bulb Temperature calculated Advanced Models Only The PosiTector DPM powers up...

Page 3: ...Bluetooth OFF Trend Chart None Perform a more thorough Hard Reset by powering down the instrument waiting several seconds then simultaneously holding both the center and buttons until the Reset symbol...

Page 4: ...uld be greater than 1 C 2 F This menu item only appears when all memory functions are turned off Wet Bulb readings cannot be stored in memory NOTE Advanced models only Displays a real time graph of th...

Page 5: ...r dimming Advanced only Memory Management Memory Management Memory The PosiTector DPM can record datasets in memory for printing to the optional Bluetooth wireless printer transferring to a computer o...

Page 6: ...the instrument will wake up take a reading sync to PosiTector net if configured and power down The instrument can record approximately 8 000 datasets while in this mode using fresh alkaline batteries...

Page 7: ...connected USB or Bluetooth to an internet connected PC running PosiTector Desktop Manager PDM Sync Now Connect Connect Menu Connect Menu The PosiTector uses a USB mass storage device class which provi...

Page 8: ...umidity sensors Keep fingers away from the sensor as body heat can cause incorrect readings Allow time for the instrument to acclimate and for the readings to stabilize when moving the instrument betw...

Page 9: ...1 F Air Temperature 40 to 80 C 40 to 175 F 0 5 C 1 F 0 1 C 0 1 F Humidity 0 to 100 3 0 1 Operating Range 40 C to 80 C 40 F to 175 F www defelsko com DeFelsko Corporation USA 2011 All Rights Reserved T...

Reviews: