12
Calibration Troubleshooting
If
Err 0
appears on the display the calibration zero is out of range. Remove all
weight from the platform and press ENTER to perform the zero calibration again.
If this error persists, consult the raw counts table below.
If
ErrSp
appears on the display the span calibration is out of range. Place the
proper weight on the platform and press ENTER to perform the span calibration
again. If this error persists, consult the raw counts table below.
The following table shows the acceptable Raw Counts for no load and full load.
Raw Counts can be viewed through the last setup parameter. Refer to Section
5, Parameter Setup, for more information regarding parameter viewing.
If the scale’s raw counts are not within the ranges specified in the following
table, contact Doran’s Technical Assistance department at 800-264-4107.
Calibration Requirements in Raw
Counts
Minimum
Maximum
Zero Calibration Point
0
50,000
Required Calibration Span
(100% of Capacity Scale Cal Point – Zero
Cal Point)
90,000
180,000
Required Calibration Span
(5% of Capacity Cal Point – Zero Cal Point)
4,500
9,000