Doran Scales DS9100 Instruction Manual Download Page 13

9

Program Wheelchair Tare Memory

1.  Press tare memory location to be changed (0 through 9)
2.  Press CHAIR RECALL.  The current tare value in memory will be displayed.
3.  Press ENTER to change the tare value in memory.
4.  Enter the tare value to be saved. Enter zero for any digit that is not required.

(For example: To enter a tare value of 41.1 LB, press 0-4-1-1)

5.  Press ENTER.

NOTE: A tare weight can be recalled when the scale is off or while a patient is
on the scale.  Recalling a tare value will initiate a reweigh function.

Recall Wheelchair Tare Manually

1.  Press CHAIR RECALL.
2.  Press memory location to be recalled. (For Example: press key 1 to recall

memory location 1)

Enter Tare Weight Manually

1.  Enter the tare weight using the numeric keypad.
2.  Press ENTER.

NOTE: A tare weight can be entered when the scale is off or while a patient is on
the scale.  Entering a new tare value will initiate a reweigh function.

Clear Tare Weight

By default, tare weights are stored in memory until cleared.  Be sure to clear any
unwanted tare weights after using the scale.

1.  Press ZERO to clear a stored tare weight

NOTE: See the

ATC

parameter in Section 5 to auto clear the tare value when the

scale turns off.

Summary of Contents for DS9100

Page 1: ...Model DS9100 Wheelchair Scale Instruction Manual Doran Scales Inc 1315 Paramount Pkwy Batavia IL 60510 USA 1 800 264 4107 www doranmedical com MAN0264 Revision 0 2...

Page 2: ......

Page 3: ...re Manually 9 Enter Tare Weight Manually 9 Clear Tare Weight 9 Section 4 Scale Calibration 10 Entering and Exiting Setup Mode 10 Calibration Procedure 10 Calibration Troubleshooting 12 Section 5 Param...

Page 4: ...unications 17 Optional RS 232 Communications Port 17 Optional USB Communications Port 17 Remote Scale Commands 17 Section 7 Specifications 19 Section 8 Troubleshooting 20 General Problem Resolution 20...

Page 5: ...e and result in costly repairs that may not be covered by the warranty Please observe the following precautions to insure years of trouble free service from your new scale Do not use sharp objects to...

Page 6: ...tilted onto the four wheels at the rear of the platform See Figure 1 Once the scale is relocated and is ready for use 1 Place the scale platform onto the floor See Figure 2 2 Pull the release lever in...

Page 7: ...e side directly opposite of the ramp Fig 3 Ramp Screw Configuration Adjusting Ramp Extension Angle The ramp is designed to not touch the floor when weight is not applied to the ramp This feature incre...

Page 8: ...Leveling the scale will provide a stable weighing surface and optimal accuracy To level the platform use the end of the flat wrench marked with an F provided with the scale Insert the wrench into pla...

Page 9: ...r to remove the screw holding the battery compartment in place Completely remove the battery compartment from the indicator and install the new set of batteries Be sure the batteries are oriented prop...

Page 10: ...form the indicator will turn on automatically The indicator has an adjustable Automatic Shutoff Timer that will power down the indicator when not in use Zero When the zero button is pressed the scale...

Page 11: ...the scale platform The Average Hold weight may be slightly less accurate than a Stable Hold weight Displaying an Average Hold weight instead of a Stable Hold weight will speed the weighment process of...

Page 12: ...m empty Note If the scale zero is not reset with an empty platform the zero weight will be retained in memory and cause weighing errors for future patients BMI Operation 1 Move the patient onto the sc...

Page 13: ...iate a reweigh function Recall Wheelchair Tare Manually 1 Press CHAIR RECALL 2 Press memory location to be recalled For Example press key 1 to recall memory location 1 Enter Tare Weight Manually 1 Ent...

Page 14: ...isplay to CAL y 3 Press ENTER 4 The display reads 000 Use the numeric keypad to enter 123 5 The current scale resolution is displayed Press ENTER 6 The display reads CAL 0 7 Remove all weight from the...

Page 15: ...ero point calibration is complete 9 The scale will display 0400 0 10 If necessary press KG to calibrate in kilograms 11 Press LAST WEIGHT 12 The scale will display 200 LB 100 kg 13 Place 200 LB 100kg...

Page 16: ...raw counts table below The following table shows the acceptable Raw Counts for no load and full load Raw Counts can be viewed through the last setup parameter Refer to Section 5 Parameter Setup for m...

Page 17: ...ll to the next parameter 2 Press CHAIR RECALL to change the current parameter setting Calibration Mode Entry Cal Enter Calibration Mode N Do not enter calibration y Enter Calibration see Scale Calibra...

Page 18: ...re memory is retained y Tare memory is cleared Automatic Shutoff Timer Ao Automatic shutoff timer The timer controls the period of time the scale stays on after non use 10 10 seconds 30 30 Seconds 60...

Page 19: ...try 0 1 0 1 increment for BMI entry 0 5 0 5 increment for BMI entry 1 1 increment for BMI entry BMI cm Entry Cnn Default parameters 0 5 0 5 cm increment for BMI entry 1 1 cm increment for BMI entry Da...

Page 20: ...d on Beep enabled of Beep disabled Stable or Average Hold Weight Beep bH Enables Disables beep when a stable or average hold weight is achieved on Beep enabled of Beep disabled Battery Voltage Level D...

Page 21: ...ing and receiving units must be set to the same baud settings The typical value is 9600 The DS6100 is factory set for eight bits no parity and one stop bit also known as 8n1 The receiving units must b...

Page 22: ...lb HEIGHTVALUE in BMIVALUE BMI CR LF Kilograms STX POL WEIGHT kg GRS TAREVALUE kg HEIGHTVALUE cm BMIVALUE BMI CR LF Data String Key STX ASCII 02 non printable character POL minus sign for negative wei...

Page 23: ...ed and lubricated Wheel Material EVA Power 6 AA Alkaline Batteries Two year battery life with 10 second Automatic Shutoff Timer setting Displayed Units lb kg Capacity 1000 x 0 1 lb 454 x 0 05 kg Const...

Page 24: ...s pressed Make sure that the scale is stabilizing Motion annunciator is off After pressing the ZERO button the scale should zero as soon as it becomes stable Weight readings are not correct Verify the...

Page 25: ...rt for assistance Errsp The span calibration is out of range This error is displayed after a span calibration attempt Place the proper weight on the platform and press BMI to perform the span calibrat...

Page 26: ...or indirect damages to equipment or personnel Nor shall Doran Scales Inc be liable for damages to equipment or for personal injury caused by misuse overload accidental damage alteration improper insta...

Reviews: