Section 3.  Scale Operation

Fig. 3 DS7100 Front Panel

Display Pushbuttons

The Model DS7100 controls consist of LAST WEIGHT, LB, KG, ZERO,
REWEIGH and BMI buttons.  The display is used to provide weight indications
and operator messages indicating scale operation.

Power On/Off

When the patient stands on the scale platform, the indicator will turn on
automatically.  The indicator has an adjustable Automatic Shutoff Timer that will
power down the indicator when not in use.

Zero

When the zero button is pressed, the scale will be reset to zero weight.  This
feature can be used to zero out any weight not desired to be part of the patient
weight.

Units Select

Press LB to weigh in pounds and enter BMI height in feet and inches.
Press KG to weigh in kilograms and enter BMI height in centimeters.
The current unit will be displayed to the right of the weight digits.

Reweigh

To reinitiate the weighing process, press REWEIGH.  This will allow a weighment
to be reinitiated without the patient having to get off the scale platform.

Summary of Contents for DS7100

Page 1: ...Model DS7100 Handrail Scale Instruction Manual Doran Scales Inc 1315 Paramount Pkwy Batavia IL 60510 USA 1 800 264 4107 www doranmedical com MAN0262 Revision 0 3 ...

Page 2: ......

Page 3: ...ation Procedure 12 Multiple Point Calibration Procedure 13 Calibration Troubleshooting 14 Section 5 Parameter Setup 15 Entering and Exiting Setup Mode 15 Calibration Mode Entry 15 Calibration Mode Selections 15 Reset All Parameters to Default Settings 15 Automatic Shutoff Timer 16 Average Aperture 16 BMI Inch Entry 16 BMI cm Entry 16 Data Output 17 Data Format 17 Baud Rate 17 LB Units Button Enabl...

Page 4: ...Remote Scale Commands 19 Section 7 Specifications 21 Section 8 Troubleshooting 22 General Problem Resolution 22 Error Messages 23 ...

Page 5: ...sult in costly repairs that may not be covered by the warranty Please observe the following precautions to insure years of trouble free service from your new scale Do not use sharp objects to press any of the buttons Do not transport patients on the scale Do not use the scale if the scale is damaged in any way Do not leave patients unattended on the scale Do not exceed the scale capacity Do not dr...

Page 6: ...3 Indicator 4 2 Handrails 5 8 Tower Bolts M6x32 6 2 Handrail Tower Bolts M6 x 16 7 2 Handrail Platform Bolts M8 x 20 8 2 Indicator Bolts M4x12 9 10 M6 Washer 10 10 M6 Spring Lock Washer 11 2 M8 Washer 12 2 M8 Spring Lock Washer 13 2 M4 Washer 14 2 M4 Spring Lock Washer 15 2 M4 Acorn Nuts 16 Foot F and Indicator I Nut Adjustment Tool 1 Carefully remove the scale from the shipping carton Be sure to ...

Page 7: ...the Platform 2 blue plastic trim caps Carefully pull the wire harness from the top of the Tower 1 as the tower is inserted into the blue plastic trim caps to prevent pinching the cable between the Tower 1 and the Scale platform 2 ...

Page 8: ...and can only be connected when inserted properly Tug gently on the cable to ensure a proper connection has been made 5 Fasten the indicator to the tower using the indicator bolts 8 washer 13 and lock washers 14 from the bottom of the Tower 1 Fasten to the top of the Tower 1 using the acorn nuts 15 The acorn nuts can be tightened with the Adjustment Tool 16 with the end marked I ...

Page 9: ...he scale backwards so that the Tower 1 is resting on the floor Insert a lock washer 10 and then a washer 9 onto the tower bolt 5 Insert each bolt but do not tighten completely until all eight bolts are installed ...

Page 10: ...ails 4 into the opening in the rubber pad Insert the handrail platform bolts 7 lock washer 11 and washer 12 but do not fully tighten Insert the handrail tower bolts 6 lock washer 9 and washer 10 into the rear of the Tower 1 Once the all handrail bolts are in place tighten all bolts fully ...

Page 11: ...location it is recommended to level the scale Leveling the scale will provide a stable weighing surface and optimal accuracy To level the platform use the Foot Adjustment Tool provided with the scale Insert the wrench into place and twist the foot to change the height while grasping the wrench After leveling the scale press Zero ...

Page 12: ...ace Completely remove the battery compartment from the indicator and install the new set of batteries Be sure the batteries are oriented properly When reinstalling the battery tray mate the key at the bottom of the battery tray to the key integral to the metal housing Gently push and hold the battery tray in place and reinstall the screw to complete the battery tray installation Fig 3 Battery Tray...

Page 13: ...ff Timer that will power down the indicator when not in use Zero When the zero button is pressed the scale will be reset to zero weight This feature can be used to zero out any weight not desired to be part of the patient weight Units Select Press LB to weigh in pounds and enter BMI height in feet and inches Press KG to weigh in kilograms and enter BMI height in centimeters The current unit will b...

Page 14: ...rm The Average Hold weight will be slightly less accurate than a Stable Hold weight Displaying an Average Hold weight instead of a Stable Hold weight will speed the weighment process of patients that cannot stand still The Average Hold weight will differ from the patient s actual weight based upon the severity of the motion caused by the patient while on the scale Although this weight will be less...

Page 15: ... memory and cause weighing errors for future patients BMI Operation 1 Step onto the scale platform 2 The display will beep when a stable or average hold weight is achieved 3 Press BMI The display will read 5 06 representing 5 feet 6 inches 4 To change the value of the flashing digit press the up and down arrows When the correct digit is displayed press BMI to advance to the next digit 5 When BMI i...

Page 16: ... The display reads 000 Use the BMI and the arrow buttons to enter 123 When the correct flashing digit is displayed press BMI to accept the current digit 5 The current scale resolution is displayed Press BMI 6 The display reads CAL 0 7 Remove all weight from the sale platform NOTE During calibration be sure vibration and air currents are not present 8 Press BMI to perform a zero point calibration T...

Page 17: ... zero point calibration The display will count down to zero The counter will reset if too much motion is detected 9 The scale will display 0400 0 10 If necessary press KG to calibrate in kilograms 11 Press LAST WEIGHT 1 The scale will display 200 LB 100 kg 2 Place 200 LB 100kg onto the scale 3 Press BMI The display will count down to zero 4 The scale will display 400 LB 200 kg 5 Place 400 LB 200kg...

Page 18: ...raw counts table below The following table shows the acceptable Raw Counts for no load and full load Raw Counts can be viewed through the last setup parameter Refer to Section 5 Parameter Setup for more information regarding parameter viewing If the scale s raw counts are not within the ranges specified in the following table contact Doran s Technical Assistance department at 800 264 4107 Calibrat...

Page 19: ...ng Calibration Mode Entry Cal Enter Calibration Mode N Do not enter calibration y Enter Calibration see Scale Calibration section Calibration Mode Selections d Select Scale Resolution lb 0 2 0 1 Default Setting NOTE This parameter is only visible when in the calibration mode CAL Zero Calibration Point 0 NOTE This parameter is only visible when in the calibration mode 0400 0 Span Calibration Point ...

Page 20: ...d a stable weight A slower setting will result in more accurate weights but will require a longer time to achieve a stable weight A faster setting will achieve a stable weight faster but will be less accurate 10 Slowest Setting 20 30 50 Default Setting 80 100 150 200 Fastest Setting BMI Inch Entry in Controls Count By of BMI entry 0 1 0 1 increment for BMI entry 0 5 0 5 increment for BMI entry 1 1...

Page 21: ... is transmitted Data Format For Format of transmitted data from the serial port FO Basic data format F1 Data collection format for use with Doran Excelerator program Baud Rate b r Serial port transmission rate 96 9600 Baud bits per second 192 19 200 Baud bits per second LB Units Button Enable Disable Serial port transmission rate on LB button active of LB button disabled KG Units Button Enable Dis...

Page 22: ...Voltage Level DC Displays current battery voltage DC 5 0 9 6 Empty Battery Symbol 5 1 5 9 V Battery Bar Indications One Bar 6 0 6 9 V Two Bars 7 0 7 9 V Three Bars 8 0 V and above Raw Counts Raw counts from A D converter View these numbers if calibration is unsuccessful Refer to Section 4 for interpretation of these numbers Press BMI to exit setup mode ...

Page 23: ...ding and receiving units must be set to the same baud settings The typical value is 9600 The DS7100 is factory set for eight bits no parity and one stop bit also known as 8n1 The receiving units must be set to 8n1 for proper communications Optional USB Communications Port The D7100 can be equipped with an optional USB communications port If this option has been installed the USB connection is loca...

Page 24: ... lb HEIGHTVALUE in BMIVALUE BMI CR LF Kilograms STX POL WEIGHT kg GRS TAREVALUE kg HEIGHTVALUE cm BMIVALUE BMI CR LF Data String Key STX ASCII 02 non printable character POL minus sign for negative weight or a space for a positive weight WEIGHT 6 character field TAREVALUE 6 character field HEIGHTVALUE 6 character field BMIVALUE 5 character field SP ASCII space CR Carriage return non printable char...

Page 25: ...UL ULc and CE approved Two year battery life with 10 second Automatic Shutoff Timer setting Displayed Units lb kg Capacity 1000 x 0 1 lb 454 x 0 05 kg Construction Aluminum Base and Column Aluminum and Plastic Housing Approvals CE 2006 95 EC Low Voltage Directive Reference Standard CENELEC EN 61010 1 2004 108 EC EMC Directive Reference Standard CENELEC EN 61326 Options RS 232 Communications USB Co...

Page 26: ...is pressed Make sure that the scale is stabilizing Motion annunciator is off After pressing the ZERO button the scale should zero as soon as it becomes stable Weight readings are not correct Verify the scale calibration with an accurate test weight If the readings are not correct recalibrate Be sure the platform is on a flat surface and all four feet are touching the floor Scale drifts off zero Ch...

Page 27: ... for assistance Errsp The span calibration is out of range This error is displayed after a span calibration attempt Place the proper weight on the platform and press BMI to perform the span calibration again Refer to the Calibration Troubleshooting section for additional information If this problem persists contact Doran Technical Support for assistance ou ld The scale is in overload The load on t...

Page 28: ...r indirect damages to equipment or personnel Nor shall Doran Scales Inc be liable for damages to equipment or for personal injury caused by misuse overload accidental damage alteration improper installation or unauthorized opening of the equipment Under no circumstances will Doran Scales Inc be responsible for any indirect or consequential damages due to errors in weighing or failure of a Doran Sc...