E8 - Diagnostics
This feature is a view-only feature. No configuration changes are possible from within this menu.
The display will alternate between the current value and diagnostic code
E:8x
: where x is one of the following variables:
0 Sensor signal, A to D counts low ppm range.
1 Sensor signal, A to D counts high ppm range.
2 Sensor signal, A to D counts %vol range.
3 Firmware version.
The diagnostic code can be selected by pressing the
[ ]
key.
Press
[ MENU ]
to return the instrument to its standard mode of operation.
E9 - Restore
This option allows the user to restore the configuration data to the factory default values. The user can restore either or both
sensor data.
The analyser must be calibrated after a restore. All previous calibration data for
the selected sensor will be lost.
Press the
[ ]
key to select the sensor type.
Press
[
]
to restore the selected sensor data. The display will show ‘
– – – –
‘ to confirm that the restore has been
performed.
Note: Pressing the [ MENU ] key rather than the
[
]
key exits the restore feature without performing any change.
Press the
[ MENU ]
key to close the menu system.
Note: The sensor zero factor will be displayed on exit while the
[ MENU ]
key is pressed.