Domel 10l User Manual Download Page 5

 

Install  the  shelf  at  the  desired  height,  with  the  shelf  edges  up.  If  you  intend  to  place  the 
samples on the bottom of the chamber, install the shelf so that it sits with the edges down on 
the  chamber  bottom,  thus  creating  an  air  gap  between  the  chamber  bottom  and  your 
samples. When installing your shelf, remember to leave clearance for the thermometer that 
will come down from the top of the chamber. 

 

Gently  insert  the  thermometer  into  the  rubberized  hole  on  the  top  of  the  incubator.  Avoid 
inserting or removing the thermometer too quickly; this can dislodge the rubber grommet that 
is used to hold the thermometer. For maximum accuracy, insert the thermometer to a point 
where the thermometer immersion line is just inside the incubator chamber. 

 

Turn the power switch to the ON position. The switch should illuminate. 

 

Set  the  temperature  control  knob  to  10.  When  the  thermometer  reaches  the  desired 
operating temperature, turn the temperature control knob down, until the heating lamp turns 
off. Wait for the temperature to stabilize. Make fine adjustments up or down as required, until 
the  desired  operating  temperature  is  obtained.  Allow  the  unit  to  stabilize  after  each 
adjustment.  Temperature  stability  is  obtained,  when  the  heating  lamp  is  turning  on  and  off 
automatically, and the temperature remains constant. 

 
 
 

6. Cleaning and Maintenance

 

 

Prior  to  any  cleaning  or  maintenance  of  the  unit,  disconnect  the  power  cord 
from the wall outlet.

 

 

No routine maintenance is required on the unit. 

 

Clean the unit internally and externally with a damp cloth and mild soap. Do not immerse the 
unit  or  spill  liquids  into  or  on  the  unit.  Use  of  alcohol  or  ammonia-based  cleaners  on  the 
window may cause it to crack. 

 

Allow the incubator to dry completely inside and outside, before reconnecting the power and 
using it again. 

 
 
 

7. Troubleshooting and Service

 

 

If the unit fails to operate, make sure that it is plugged into a properly grounded working wall 
outlet.  Check  that  the  unit  is  switched  on,  the  power  switch  is  illuminated,  and  that  the 
temperature control knob is turned up. If necessary, check the fuses. 
If the problem persists, contact your distributor.

 

 

Summary of Contents for 10l

Page 1: ...elezniki Slovenia BU Laboratory Systems Na Plavžu 79 4228 Železniki Slovenia Tel 386 0 4 5117 500 Fax 386 0 4 5117 501 Internet http labs domel com ANALOGUE INCUBATOR 10L USER S MANUAL Edition Month Year 3 October 2019 ID 421836 ...

Page 2: ...ing 3 3 Installation 3 3 1 Power Source 3 3 2 Location 4 4 Control panel 4 4 1 Power Switch 4 4 2 Heating Lamp 4 4 3 Temperature Control Knob 4 5 Operation 4 6 Cleaning and Maintenance 5 7 Troubleshooting and Service 5 EU Declaration of Conformity 6 ...

Page 3: ...pect damage to the unit immediately file a claim with the carrier in accordance with their instructions before contacting your distributor Also verify that all accessories are included and that the unit is in good working order before discarding the shipping packaging Analogue Incubator 10l is shipped with the following Stainless steel shelf Thermometer Power cord 3 Installation Your satisfaction ...

Page 4: ...en the unit is energized 4 2 Heating Lamp This light is ON when the unit is heating When the unit has reached the set temperature the light will be on only when the unit is heating to maintain the temperature 4 3 Temperature Control Knob This control knob is marked with temperature symbol It controls the incubator chamber temperature on a scale from 0 to 10 5 Operation Do not use this incubator in...

Page 5: ... the heating lamp turns off Wait for the temperature to stabilize Make fine adjustments up or down as required until the desired operating temperature is obtained Allow the unit to stabilize after each adjustment Temperature stability is obtained when the heating lamp is turning on and off automatically and the temperature remains constant 6 Cleaning and Maintenance Prior to any cleaning or mainte...

Page 6: ... use within certain voltage limits Standards EN 61010 1 2010 EN 61010 2 010 2014 2 Directive 2014 30 EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to electromagnetic compatibility Standards EN 61326 1 2013 3 Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of th...

Reviews: