
5
Install the shelf at the desired height, with the shelf edges up. If you intend to place the
samples on the bottom of the chamber, install the shelf so that it sits with the edges down on
the chamber bottom, thus creating an air gap between the chamber bottom and your
samples. When installing your shelf, remember to leave clearance for the thermometer that
will come down from the top of the chamber.
Gently insert the thermometer into the rubberized hole on the top of the incubator. Avoid
inserting or removing the thermometer too quickly; this can dislodge the rubber grommet that
is used to hold the thermometer. For maximum accuracy, insert the thermometer to a point
where the thermometer immersion line is just inside the incubator chamber.
Turn the power switch to the ON position. The switch should illuminate.
Set the temperature control knob to 10. When the thermometer reaches the desired
operating temperature, turn the temperature control knob down, until the heating lamp turns
off. Wait for the temperature to stabilize. Make fine adjustments up or down as required, until
the desired operating temperature is obtained. Allow the unit to stabilize after each
adjustment. Temperature stability is obtained, when the heating lamp is turning on and off
automatically, and the temperature remains constant.
6. Cleaning and Maintenance
Prior to any cleaning or maintenance of the unit, disconnect the power cord
from the wall outlet.
No routine maintenance is required on the unit.
Clean the unit internally and externally with a damp cloth and mild soap. Do not immerse the
unit or spill liquids into or on the unit. Use of alcohol or ammonia-based cleaners on the
window may cause it to crack.
Allow the incubator to dry completely inside and outside, before reconnecting the power and
using it again.
7. Troubleshooting and Service
If the unit fails to operate, make sure that it is plugged into a properly grounded working wall
outlet. Check that the unit is switched on, the power switch is illuminated, and that the
temperature control knob is turned up. If necessary, check the fuses.
If the problem persists, contact your distributor.