3
1. Specifications
Analogue Incubator 10l is a compact general-purpose convection incubator capable of
uniformly heating samples up to a maximum of 60
°C. The unit comes with a movable shelf
and has a door window area, so samples can be observed without opening the door, thus
preventing heat loss.
Temperature range:
A5
°C to 60 °C
Controls:
Analogue - hydraulic
Unit size W x D x H:
28.5 x 28 x 33.5 cm
Chamber W x D x H:
23 x 22 x 20 cm
Unit weight:
8.3 kg
Electrical:
230 V, 50/60 Hz, 0.3 A or 115 V, 50/60 Hz, 0.6 A
Fuses:
2 x 0.5 AT (230 V) or 2 x 1 AT (115 V)
2. Unpacking
Carefully inspect the unit and accessories upon receipt for any damage that may have
occurred during shipping. Severe damage to the packaging container may indicate damage
to the unit. If you suspect damage to the unit, immediately file a claim with the carrier in
accordance with their instructions before contacting your distributor. Also verify that all
accessories are included, and that the unit is in good working order, before discarding the
shipping packaging.
Analogue Incubator 10l is shipped with the following:
• Stainless steel shelf
• Thermometer
• Power cord
3. Installation
Your satisfaction and safety require a complete understanding of this unit.
Read the instructions thoroughly and be sure all operators are given adequate
training before attempting to put the unit in use.
NOTE: This equipment must be used only for its intended application; any alterations
or modifications will void your warranty and may cause injury.
3.1 Power Source
Check the data plate for voltage, cycle, wattage and ampere requirements. If matched to
your power source, plug the power cord into a grounded outlet. Voltage should not vary more
than 10% from the data plate rating.