26 / 27
DRCA 1-CT transducer
Rated current
DRCA 1-CT 70
200 A
DRCA 1-CT 105
250 A
DRCA 1-CT 140
350 A
DRCA 1 measuring unit
Design requirements
The device was manufactured
in accordance with DIN EN 61010-1
(VDE 0411 Part 1): 2002-08
'Safety requirements for electrical
equipment for measurement,
control and laboratory use'.
Note
����������������
With the rated currents stated above, a maximum measurement error
of 10 mA can be expected with a centred line structure for the primary
load current. This error is the result of asymmetrical currents within the
transformer core material.
DRCA 1-SW software
Operating system
Windows 7 SP1 or higher
Memory requirements
500 MB free hard disk storage
10.1
Product overview/Article numbers
DRCA 1 set
09 352 050
consisting of:
DRCA 1
09 352 051
DRCA 1-CT 70
09 352 052
DRCA 1-MC
09 352 053
DRCA 1-SW
09 352 054
10. Technical data
DRCA 1 measuring unit
Rated voltage
+5 V (USB)
Wiring
USB cable
Sampling rate
250 kHz
Damping
40 dB/decade greater than 100 kHz
Frequency range:
10 Hz to 100 kHz
Measuring range
AC: up to 10 A peak-to-peak value,
DC: not measurable
Accuracy when I = 100 mA
when residual current I = 100 mA
10 Hz - 100 kHz: ± 1 dB
20 Hz - 20 kHz: ± 0.3 dB
1 kHz: ± 0.1 dB
Resolution
0.2 mA
Pollution degree
2 (for indoor use only)
Operating temperature range 0 °C to 45 °C
Storage temperature range
-20°C to 70°C
Resistance to climatic changes
max. 90% rel. humidity,
condensation not permitted
Operating altitude
max. 2000 m above sea level
Dimensions
250 mm × 160 mm × 40 mm
(L × W x D)
Weight
0.85 kg
Positioning
optional
Measurement category
I
(please note: the device must
not be used in measurement
categories II, III and IV)
TECHNICAL DATA