background image

Operation Manual 

 

 

VacuDAP 2002

 

2002BA01e.DOC / 22.07.2004

 

 

20

 

of

 

22

 

When doing this, the following requirements should be met: 

• 

Adjust the field size of the useful beam at a suitable focal distance (e.g. X-ray table) to 
approx. 10 cm x 10 cm. 

• 

Expose the X-ray film (without a supporting foil, where possible) to the radiation at this 
focal distance (high contrast, optical density 

 0.5). 

• 

Measure the dose (air kerma, 

'

K

) in the centre of this field at an uncertainty of 

 5% 

(keep the influence of back scattering as low as possible). 

• 

Set the X-ray equipment to 70 kV; (20-100) mA. 

• 

Make the test with the VacuDAP 2002 measuring system completely installed. 

 

 

If measurements are made both with and without an absorber (see 
2.7,

 Factory calibration

), the test and, where necessary, re-calibration 

must be made separately for the two calibration factors.  

 

Provided the above requirements are satisfied, the following test procedure can be applied:  

1.  Measure the room temperature 

T

 and the atmospheric pressure 

p

.

 

2.  Find the correction factor for the air density 

)

,

(

p

T

f

k

=

ρ

 (see data sheet). 

3.  Expose and process the X-ray film, and accurately measure the field edge lengths 

2

1

;

l

l

4.  Measure the dose 

'

K

 

in the same plane as in the previous exposure of the film, and 

simultaneously determine the dose area product 

A

K

 using the installed VacuDAP 2002 

measuring system. 

5.  Calculate the reference value for the dose area product 

'

'

A

K

 from: 

 

 

2

1

'

'

'

l

l

K

A

K

=

 ; 

where 

'

K

....

...dose (air kerma) in mGy 

 

 

 

 

 

  

2

1

;

l

l

….edge length of exposed area in cm 

6.  Determine the deviation,

F

, of the indicated value 

A

K

 from the reference value 

'

'

A

K

'

'

)

(

A

K

A

K

k

F

=

ρ

 

 

ρ

k

........correction factor (see data sheet) 

If the deviation does not exceed ±10%, i.e.:  

1

.

1

9

.

0

F

, a re-calibration is not necessary. 

If the deviation is greater, calculate the new calibration factor, 

CF_ABOVE

 new

 (or 

CF_UNDER

 new

) as 

follows: 

 

 

F

old

new

 

CF_ABOVE

 

CF_ABOVE

=

 

The old calibration factor 

CF_ABOVE

 old

 (or 

CF_UNDER

 old

) can be read from the display, when the 

display unit is in the parameter changing menue (

SETTINGS

, see 3.3.9, 

Changing of parameters

) at 

the appropriate step (

CALIB

CF_ABOVE

, see fig. 5, 

Changing settings

). If necessary, the new 

calibration factor can be entered there as well. 

 

After the calibration factor has been changed, the calibration must be re-
checked. 

Summary of Contents for VacuDAP 2002

Page 1: ...ystem VacuDAP 2002 valid from serial number 03 00001 valid from software version 9 5 VacuTec Messtechnik GmbH Tel 49 0 351 3 17 24 0 Dornbl thstra e 14 Fax 49 0 351 3 10 50 85 D 01277 Dresden Email in...

Page 2: ...a parameter that comes in very useful when measuring the radiation doses patients are exposed to in a radiological examination The VacuDAP 2002 comprises a circular ionization chamber miniaturised ele...

Page 3: ...ration 8 3 Operating instructions 9 3 1 General 9 3 2 Components and measuring principle 9 3 2 1 Ionization chamber and detector electronics 9 3 2 2 Display units connections and control elements 10 3...

Page 4: ...dard limit values to IEC 55011 group 1 class A and satisfies the EMC protection requirements as specified in the law on medicinal products 1 4 Protection against electric shock At the power supply end...

Page 5: ...a cloth soaked in water with a mild washing up liquid added is possible While cleaning make sure no moisture gets into the ionization chamber The chamber must be dry before starting the system again...

Page 6: ...ent carrying capacity of 300 mA in case of a 2 channel measuring system 956 00 01 500 mA Further requirements can be found in Chapter 1 4 Protection against electric shock The supply voltage is applie...

Page 7: ...test is started If the VacuDAP 2002 measuring system was correctly installed the supply voltage can be applied After an initialisation phase the measuring system starts the POWER ON TEST see 3 3 3 TE...

Page 8: ...ated together by the manufacturer A re calibration is necessary if ionization chamber and detector electronics with different serial numbers are used together see 4 4 Checking the overall calibration...

Page 9: ...ring principle Fig 1 Components of the VacuDAP 2002 measuring system Within the ionization chamber 1 the X rays produce charge carriers which are proportional to the product of dose and X rayed ioniza...

Page 10: ...r or PC The measured values in mGy cm are displayed by a high intensity 8 digit LED display The readout resolution is 1 mGy cm The measuring system is operated by using the two keys Print and Reset on...

Page 11: ...3 3 1 Schematic diagram Figure 2 shows the program flowchart of the measuring system The various functions will be described in detail in the following sections Print Reset Reset Print Reset Print Res...

Page 12: ...amber are tested under measuring conditions A TESTACT value is determined and compared with a TESTNOM value The TESTNOM value is ascertained in the course of the factory calibration stored as a system...

Page 13: ...e is displayed on the display unit For example a zero is displayed upon the correct completion of the TEST function 0 Upon activation of the MEASURE function the system starts to measure all X rays in...

Page 14: ...ASCI characters to a PC To do this parameter PRINTER must be set accordingly see Table 4 A special interface cable is required for the purpose see 6 Accessories and options The type of data printout...

Page 15: ...ered in radiographic and fluoroscopic applications the VacuDAP 2002 measuring system allows two measuring modes to be set see Table 6 Display Measuring mode Readout during RESET function RADIOGRA Radi...

Page 16: ...rtificate TEST_VAL 2751 50 9999 TESTNOM value see calibration certificate POSITION ABOVE ABOVE UNDER Position of measuring chamber CALIB Changing calibration CF_ABOVE 1 00 0 50 1 75 Calibration factor...

Page 17: ...changed example entering a new test value TEST_VAL Parameter TEST_VAL is required to be changed when the ionization chamber of the measuring system needs to be replaced e g for repair work POSITION P...

Page 18: ...Print Print Print Print Print Print Reset Reset Reset Reset Reset Reset Reset Reset Reset Reset Reset Reset Reset Reset CF UNDER CAL OK CF ABOVE TEST VAL CHAMBER PRINTER ABOVE LOAD SAVE UNDER LOADED...

Page 19: ...down the indicated value 4 Leave the measuring system in the MEASURE mode for a rest period of at least one hour without exposing it to radiation 5 After this time read the value a second time and co...

Page 20: ...ensity p T f k see data sheet 3 Expose and process the X ray film and accurately measure the field edge lengths 2 1 l l 4 Measure the dose K in the same plane as in the previous exposure of the film a...

Page 21: ...must be repaired ZC ERROR Reset was pressed during exposure Chamber leakage current too high Repeat Reset without irradiation Repeat RESET function after drying the measuring chamber at max 50 C Meas...

Page 22: ...nter Typ DPU 414 Printer cable SEIKO 952 00 40 Standard length 2 5 m max 15 m possible Printer EPSON 950 00 03 Dot matrix printer Typ TM U210D Printer cable EPSON 952 00 44 Standard length 2 5 m max 1...

Reviews: