DMT v3.2 User Manual A-1
DMT v3.2
Page 11 (55)
2015-10-13
Open Distribution
4.1
Screen
Measurement Dashboard
Selecting which Screen measurement to view
•
Select “Screen” in the “
Type”
option (top left corner)
•
Select the “
Date
” when the measurement was done
•
Scroll down to the time of the measurement in the “
Date
/
Time
” window.
The Dashboard Content:
Result:
Shows the overall result of the measurement (provided by the
instrument’s internal algorithms). There are several options here;
•
Regulated Substance
indicates that the substance found is regulated
for the “Scan Type(s)” selected.
•
Masking Agent
indicates that the substance found may be used to
mask/hide the presence of a regulated substance for the “Scan
Type(s)” selected.
•
Precursor
indicates that the substance found is classified as a
precursor for the “Scan Type(s)” selected.
•
Adulterant
indicates that the substance found is classified as a
“cutting agent” (or adulterant) for the “Scan Type(s)” selected.
•
Clear
indicates that the substance found is not regulated for the “Scan
Type(s)” selected.
•
Inconclusive
indicates that there is no spectra in the selected
libraries that match the sample.
•
Not Categorized
indicates that the substance found has not yet been