PD030 DIODE LASER MANUAL

4. Operation Instructions continued.

4.2 Self-Detecting

a.  

When the machine is switched on, only wavelength and notice area are activated,  

 

the other functions are locked(shown in gray).

b.  

Please watch and follow the information shown in the notice area.

1)  

Connect the fiber into the fiber hole, the notice area will show “SM905 connected” if  

 

the fiber is connected correctly.

2)  

Please wait for the system to recognize the fiber diameter. The optical fiber area will  

 

be activated if the fiber is recognized.

3)  

Insert the fiber pen into the power detection hole, then the system starts power-on  

 

self-calibration. When self-calibration is finished, all functions will be activated.

c.  

Take out the fiber pen according to the notice.

4.3 Parameter Setting

a.  

Power setting

Press “Power”, the “Key Parameter” area will be highlighted, press “   ” “    ” to adjust power, 

then press “Enter” to confirm the setting.

b. Emitting mode and detail parameters setting

There are two alternative emitting modes: continuous pulse and repeat pulse.

The default mode is continuous. The waveform schematic diagram is shown in “Time 

parameter” area. If you want to choose repeat pulse, press “Repeat Pulse” in “Mode” area, 

the waveform, idle time, duty time will be shown in “Time parameter” area. Press the 

parameter you want to adjust and adjust by press number buttons or “+” “-”. Press “Enter” 

to confirm.

Key Parameter Area

Mode Area

Summary of Contents for PD030

Page 1: ...www dlc com au For further information please call 1300 785 401 Northern Office 1300 785 405 Southern Office Email sales dlc com au www dlc com au Model PD030 30W Diode Laser Surgical System Manual Please read the manual carefully before operating and keep this manual for future reference ...

Page 2: ...6 Fiber 6 2 Components 6 3 Pre startup preparations 7 3 1 Unpacking and inspection 7 3 2 Checking power voltage 7 3 3 Connecting power cable 7 3 4 Footswitch Connection 7 4 Operation Instruction 8 4 1 Power on 8 4 2 Self detecting 9 4 3 Parameter Setting 9 4 4 Change Status and Emit 10 4 5 Automatic Calibration 11 5 Fiber Cutting 12 6 Protection 12 6 1 Overheat Protection 12 7 Precautions 13 8 Mai...

Page 3: ...General requirements for safety IEC 60601 1 2 Medical electric equipment general requirement for safety collateral standard electromagnetic compatibility requirement and test IEC 60601 1 4 Medical electric equipment general requirement for safety collateral standard programmable medical electric equipment IEC 60602 2 22 Medical electric equipment part 2 specific safety requirement on diagnosing an...

Page 4: ...tle pain and few complications on the patient and the hospitalization is short The laser can match all kinds of endoscopies by using optical fiber of different diameters The laser also has flexible application in open surgery The instrument can be applied to general surgery gynecology otolaryngology dermatology and cosmetology etc for different treatment such as cutting vaporizing cauterizing and ...

Page 5: ...trument based on a diode laser module providing up to 30W output power on body tissue It is easy and safe to operate 1 3 Main cabinet 1 Diode laser and aiming beam 2 Power supply 3 Main control panel LCD touch screen 4 Power meter 5 Foot switch 6 Fiber 1 3 1 Diode laser and aiming beam Diode laser emission from the diode module when power on the forward voltage excited electronic transition laser ...

Page 6: ...ut When the footswitch is pressed the power is on and laser emits from the diode module 1 3 6 Fiber The fiber is connected to the system by sm905 the system can identify fiber with diameter of 200um 400um 600um 800um The fiber can be used in a fiber pen or with an endoscope 2 Components 1 Fiber pen shelf 2 Emergency stop switch 3 LCD touch screen 4 Power measure hole 5 Key switch 6 Fiber hole 7 Fi...

Page 7: ...t 3 2 Checking power voltage Ensure that the power voltage complies with the requirement of the instrument Socket is in working order 3 3 Connecting power cable Plug the two terminals of the power cable into the power input socket and the power socket Make sure the ground socket is in working order 3 4 Footswitch Connection Plug the footswitch cable into the socket on the rear part of the instrume...

Page 8: ...machine b Turn on the power switch at the backside then wait for 30 seconds c Insert the key into the key switch turn clockwise to power on The logo as shown in figure 1 will be displayed for 3 seconds then switches to user interface Caution After turn on the power switch you should wait for 30 seconds before turn on the key switch Figure 1 User Interface ...

Page 9: ... hole then the system starts power on self calibration When self calibration is finished all functions will be activated c Take out the fiber pen according to the notice 4 3 Parameter Setting a Power setting Press Power the Key Parameter area will be highlighted press to adjust power then press Enter to confirm the setting b Emitting mode and detail parameters setting There are two alternative emi...

Page 10: ...he treatment time When the footswitch is pressed the laser emits for the time you set Caution Pre set pulse and Pre set time are designed for experienced users We strongly recommend that new users use free output Press the output mode you want and set the value same as detail parameter setting then press Enter to confirm 4 4 Change Status and Emit After all parameters are set and confirmed press R...

Page 11: ...et up the power then insert the fiber pen into the power meter hole the system would change to auto calibration interface 2 3 Press OK to start the calibration 4 Please operate according to the notice until the calibration is finished then take out the fiber pen Caution Do not take out the fiber pen when calibration is in process When calibrating the areas are locked except the notice area ...

Page 12: ...ert The fiber into the hole hold strippers tightly on both sides pull it remove two layers of skin if the skin is not totally stripped strip again 4 Open the Fiber scissors and put fiber in gently cut into the blade around the optical fiber then cut the fiber 5 If the aiming beam displays a complete circle on the paper then cut success otherwise repeat step 4 6 Insert the fiber into fiber pen 6 Pr...

Page 13: ...ectly at the laser operators must use safety eye wear when operating 5 This instrument generates high voltages inside No attempt should be made by non professional to open the cabinet of the instrument to avoid electric shock risk 6 If the instrument gives out abnormal smell or sound stop operation at once Cut off the power first before any inspection 7 Keep the instrument in an environment with t...

Page 14: ...r wet cloth in case the water leaks into the inner part of the instrument causing short circuit and damage 8 2 Power calibration The practical laser output power and the preset panel power must be calibrated each year with standard laser power meter within validity period by trained of professional personnel 8 3 Fuse replacement Open the fuse holder with a screwdriver and remove the original fuse ...

Page 15: ... Technical Specifications Working Modes Continuous Repeat pulse Output Power 0 30W adjustable Laser Type Diode laser Wavelength 980nm Duty Time 0 01 1s Idle Time 0 01 1s Aiming Beam 635nm diode laser Delivery System 200um 400um 600um 800um diameter fiber Control System Foot switch LCD touch screen Temperature 5 40 Relative Humidity 80 Power Input AC220V 10 50Hz Weight 10kg Dimensions 43 30 20cm ...

Page 16: ... is properly used Any damage cause by improper use of the instrument such as using unfitted power supply and wrong accessories operating in a manner other than specified in this operators manual damages caused by transportation accidents unauthorized installation or maintenance etc such free service will be invalid immediately The free service does not include accessories transportation free and d...

Page 17: ...n 1 Warning and explanation label 2 Warning label Hazard symbol 3 Emergency stop switch 4 Position 1 is power ON power connection to the mains Position 0 is power OFF power disconnection from the mains 5 Attention consult ACCOMPANYING DOCUMENTS 6 Aperture label 7 HIGH VOLTAGE Dangerous label 8 Protect earth ...

Page 18: ... 405 Email info dlc com au PO Box 4438 17 19 Horne Street Hoppers Crossing VIC 3029 Northern Office Ph 07 5428 3666 Fax 07 5428 3777 Toll Free 1300 785 401 Email qld dlc com au PO Box 833 Unit 6 1 Lear Jet Drive Caboolture QLD 4510 www dlc com au www dlcimaging com au ...

Reviews: