High Speed (Refrigerated) Micro-Centrifuge
23
0.2ml PCR tube
A02P2
15000 6.9 17350
0.5ml micro
tube
A05P2
15000
7.6 19100
2 AS36-05 0.5ml
micro
tube
15000
8.5 21380
PCR8 serial
tube
A02P05
15000 7.6 19100
3 AS4-PCR8 PCR8
serial
tube
6.5/7.2
16350/18100
11.1.3 Notice
(1) The centrifuge rotor can separate sample which density lower than 2.0g/ml, if the samples density
is over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
(
2
)
To prevent corrosion, remove the rotor from rotor chamber if do not use for a lone term, then
detach the rotor lid and upside the rotor down to dry the tube holes.
(
3
)
If some samples leaked in the rotor hole, wash the hole with water, apply a thin coat of silicon
grease on the rotor surface after drying.
(
4
)
It is necessary for a regular maintenance for rotor, recommend to clean it each 3 months to keep
cleaning of tube hole and shaft hole, and then apply a thin coat of silicon grease on it.
11.1.4 Autoclaving
121
℃
(
1.0kg/cm
2
)
20 minutes. The rotor makes by high-strength aluminum alloy material, can be
high-pressure sterilization: 121
℃
(
1.0kg/cm
2
)
20 minutes.
11.1.5 Bio-safe
seal ring
The rotor is sealed by bio-safe structures, achieved using three high-temperature rubber seal rings. The seal
rings may fall off or aging after several autoclaving, need to be replaced or re-installed. The replacement
methods please refer to the section 9.3.
11.2 Tubes
11.2.1 Clean and sterilize tubes, please refer to the table below.
Table 11.2 Cleaning and sterilizing conditions for tubes
O
:
Applicable X
:
Inapplicable
Condition Material
PA
PC
PP
Clea
n
i
Acidic
(
pH5 or lower
)
X X X
Acidic
(
higher than pH5
)
O O O