DLAB Laboratory Instruments D3024 User Manual Download Page 20

  High Speed (Refrigerated) Micro-Centrifuge 

 

 

19

detach the rotor door and upside the rotor down to dry the tube holes. 

„

 

If a sample is leaked in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to 

the rotor when it is completely dried. 

„

 

The rotor should be regular maintenance, recommend to cleaning it each 3 months to ensure tube 

and rotor holes keep clean, and then apply a thin coat of silicon grease. 

6

Drain (D3024R) 

„

 

The centrifuge is equipped with 

drainage

, drain off water when more water is in drain pipe. 

 

9.2 Consumables     

The table list below is the centrifuge consumables. It is recommended to replace consumables 

according the table below. 

 

No. Consumables 

Replacement 

conditions 

Seal ring of door   

Crack 

Rubber block of temperature sensor   

Seal ring of centrifuge chamber (3024R) 

 

9.3 The replacement of seal rings   

9.3.1 Instructions 

There are three high-temperature rubber seal rings that equipped into rotor to achieve bio-safe. The seal 
rings may fall off or aging after several autoclaving, need to be replaced or re-installed. 

                   

 

 
 
                                        Figure 9-1      Seal rings of rotor 
 

9.3.2 Replacement methods 

1

Clean the seal ring slot with neutral detergent solution and make it dry.   

2

Evenly coated with glue (501) in the seal ring slot and keep the seal ring into slot, press evenly to 

Seal ring

 1  2    3 

Seal ring

 2    3 

Summary of Contents for D3024

Page 1: ...R High Speed Refrigerated Micro Centrifuge Dragon Laboratory Instruments Limited www dragon lab com Before using this centrifuge please carefully read this user manual for its efficient operation and...

Page 2: ...reparation 12 7 1 Prepare the sample 12 7 2 Inject the sample into the tubes 12 7 3 Keep the tubes balance 12 7 4 Inspect the rotor 12 7 5 Symmetrically load centrifuge tubes in rotor 13 8 Operation 1...

Page 3: ...ntrifuge 26 14 2 Guarantee of the rotor 26 After sales service 26 Copyright No part of this manual may be reproduced or transmitted without prior written permission of DRAGONLAB We can not be responsi...

Page 4: ...WARNING Personal Danger Warning notes indicate any condition or practice which if not strictly observed could result in personal injury or possible death CAUTION Possible damage to instrument Caution...

Page 5: ...advance and then notice the service center involved in the details of the particular materials z Do not handle the power cord or turn on or off the POWER switch with wet hands to void electrical shoc...

Page 6: ...e allowable minimum speed of the rotor kits rotor bucket or the other adapter Make sure to run it below the allowable minimum speed 3 Do not exceed the allowable imbalance 4 Use the rotor and tubes wi...

Page 7: ...z 60Hz 5A D3024R Single phase 220V 240V 50Hz 60Hz 10A 110V 120V 50Hz 60Hz 10A Dimensions mm D3024 L 280 W 364 H 266 D3024R L 338 W 580 H 324 Weight D3024 12kg D3024R 30kg Additional features Speed RCF...

Page 8: ...to ensure your safety and the optimum performance Before moving the centrifuge the rotor must be removed WARNING Incorrect power supply may damage centrifuge Make sure the power source conforms to the...

Page 9: ...grounded properly 1 This centrifuge is equipped with a 3P flat plug grounding can be done just by plugging the 3P plug into the outlet 2 An over 10A outlet which has a good ground is required and meet...

Page 10: ...rated Micro Centrifuge 9 Figure 5 2 D3024 Rear view of centrifuge D3024R Figure 5 3 D3024R Front view of centrifuge Outlet Seal ring of door Operation panel Door lock hook POWER switch Cover Outlet Ai...

Page 11: ...ed Refrigerated Micro Centrifuge 10 Figure 5 4 D3024R Rear view of centrifuge 6 Operation panel Air vents View Door 1 2 4 3 Speed display area Temperature display area Time display area 5 15000 25 12...

Page 12: ...the button during centrifugation 5 Parameter button Clockwise rotate to increase parameter values counter clockwise rotate to decrease parameter values Press down the button shift between speed and RC...

Page 13: ...ol displays the ratio of working to time setting The total time setting is divided into 10 scales 7 Rotor Preparation 7 1 Prepare the sample 7 2 Inject the sample into tubes CAUTION Do not inject over...

Page 14: ...the rotor may be moved off while rotating and cause damage of the centrifuge or rotor z Firmly tighten the rotor door with rotor 8 Operation 8 1 Normal operation Turn on the POWER switch centrifuge w...

Page 15: ...figure indicates the instrument has accumulated running time 312 hours 56 minutes and 45 seconds and then the instrument displays the last running parameters Figure 8 3 Last running interface Speed 15...

Page 16: ...tor and ensure is in position Close the door and then start running The method of removing rotor is as same as the above mentioned by turning the screw counterclockwise 8 1 2 Set the operation paramet...

Page 17: ...3 Start the operation 1 Press button to start running The door should be locked before rotor starts rotating It will start timing once the rotor starts running the screen displays remaining run time 2...

Page 18: ...eration 1 Turn on the POWER switch 2 Set a RCF value CAUTION Do not exceed the allowable maximum RCF value of the rotor and adapters Press the select button choose speed unit g the speed symbol will f...

Page 19: ...t with a cloth or sponge moistened with a neutral detergent solution Sterilize the centrifuge by wiping with a cloth moistened with 70 ethanol solution 2 Rotor chamber Once the rotor chamber is dirty...

Page 20: ...below is the centrifuge consumables It is recommended to replace consumables according the table below No Consumables Replacement conditions 1 Seal ring of door Crack 2 Rubber block of temperature sen...

Page 21: ...it is not available to operate the alarm code will be displayed on the time display screen and the operator can know the malfunction caused by the alarm code WARNING Do not open the door when the roto...

Page 22: ...lly once the operation finished 3 Press button once the rotor stops door can be released 10 2 2 In the case of power outage If the door cannot be opened due to the power outage open the door as follow...

Page 23: ...tor structure 11 1 2 Available rotors and adapters All rotors are used for bio safe when the rotor lid was tightened with the rotor centrifuge tubes will be enclosed into rotor to ensure the sample do...

Page 24: ...ly a thin coat of silicon grease on the rotor surface after drying 4 It is necessary for a regular maintenance for rotor recommend to clean it each 3 months to keep cleaning of tube hole and shaft hol...

Page 25: ...9 11 2 3 Autoclaving PA PC and PP tubes PA begins softening at about 120 PC and PP at about 130 Autoclave PA tubes at 115 0 7kg cm2 for 30 minutes and PC and PP tubes at 121 0 1kg cm2 for 20 minutes...

Page 26: ...High Speed Micro Centrifuge with AS24 2 rotor kits Euro plug 220V 50Hz 60Hz 912315120000 D3024 High Speed Micro Centrifuge with AS24 2 rotor kits UK plug 220V 50Hz 60Hz 922015130000 D3024R High Speed...

Page 27: ...even if within the guarantee period expires 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by conveyance or relocation after installation...

Page 28: ...fuge 27 E mail info dragon lab com Service service dragon lab com Website Http www dragon lab com US office DRAGONLAB LLC 500 Four Rod Road Suite 122 Berlin CT 06037 USA Tel 1 860 828 5289 Fax 1 860 8...