9
Intelect Advanced
Battery Module
BATTERY MODULE INSTALLATION
Installing Battery Module to System (continued)
3. Position the System over the
Module. The Ribbon Cable
must be positioned as illustrated.
4. Set the System onto the Module
making certain the feet of the
System rest in the indents in the
Module.
RIBBON
CABLE
MUST
BE AS
SHOWN
7. Set the assembly upright on the
work surface.
8. Install the Lanyard to the new
Extended Front Access Panel.
5. With System positioned on the
Module, lay the assembly on
one side.
6. Secure the Module to the
System with the four 4mm X
20mm screws using the #1
Phillips screwdriver.
NOTE:
Tighten screws until System will
not move on the Module.
9. Route the power cord through
the Battery Module to the
System.
NOTE:
If the System is equipped with the
optional Therapy System Cart, refer
to the Intelect Therapy System Cart
User Manual for proper installation
of the System with Module to Cart
instructions.
CAUTION
If Ribbon Cable is twisted, the pins will not properly align. If power is
applied to the system with misalignment of pins or a twisted ribbon
cable, the controlling electronics in the Battery Module will be destroyed
and possible damage to the system internal components could occur.