background image

805-0975-650

23

TM/MC

REORDER

PART NUMBERS

Fresh gas elbow

850-0030-000

Quick Connect

850-LC10004

Summary of Contents for Moduflex Access 2

Page 1: ...Printed in Canada ...

Page 2: ...m TABLE OF CONTENTS User Responsibility 3 Standard and optional features 4 Receiving and Assembly Instructions 5 10 Optional accessories Assembly Instructions 12 Daily Leak Check Procedure 17 18 Annual Maintenance 19 Consumables Materials List 20 Warranty 21 Order Part Numbers 22 23 ...

Page 3: ...be replaced immediately Should a repair or replacement become necessary Dispomed recommends that a request be placed by phone or in writing to the nearest Dispomed regional service center This product or any of its parts should not be repaired other than in accordance with written instructions provided by Dispomed and by Dispomed trained personnel The product may not be altered without prior writt...

Page 4: ...of leaks Quick connect fresh gas outlet for easy and secure switching between breathing circuits Cage mount compatible manifold for easy and quick installation of vaporizer Five caster spider base 1500 ml CO2 absorber designed for easy and fast replenishing of sodalime Condensation trap to prevent water build up in sodalime Easy and efficient gas sealing Patient circuit with 2 and 3 liter bags and...

Page 5: ...n components have been detached from the unit These components are packed separately within the shipping crate INSPECTION On receipt of your unit inspect the shipping container and the enclosed components for signs of damage Check your Materials List to ensure that all the necessary components are included RECEIVING AND ASSEMBLY INSTRUCTIONS x1 x1 x1 x1 x1 x1 x1 x1 x1 x1 x1 ...

Page 6: ... MC DISPOMED Ltd 1 800 363 1746 info dispomed com www dispomed com 1 RECEIVING ASSEMBLY INSTRUCTIONS CONTINUED Remove the accessories found in the bag inside the large box Remove the wheeled base and post from the box 2 ...

Page 7: ... 0975 650 7 TM MC 4 Remove the head assembly from the inner box and slide it onto the top of the black post RECEIVING ASSEMBLY INSTRUCTIONS CONTINUED 3 Slide the tapered end of the black post into the base ...

Page 8: ...omed com www dispomed com 5 6 Install the vaporizer as indicated on page 12 Install a breathing bag on the rotary bag mount of the absorber RECEIVING ASSEMBLY INSTRUCTIONS CONTINUED Tighten the head assembly on the back of the head using the Allen key supplied ...

Page 9: ... the front of the absorber Fill the canister with absorbant using pre packed or loose fill granules caution do not overfill the canister leave approximaly 1 4 5mm from the top empty Install the canister and tighten the sealing screw RECEIVING ASSEMBLY INSTRUCTIONS CONTINUED 7 ...

Page 10: ...om 9 10 Install the oxygen supply hose to the back of the unit using 2 wrenches 5 8 and 11 16 in order to reduce the risk of accidentally disconnecting the hose Connect the other end of the oxygen supply hose to a 50 p s i 3 bar oxygen supply RECEIVING ASSEMBLY INSTRUCTIONS CONTINUED ...

Page 11: ...805 0975 650 11 TM MC ASSEMBLED VIEWS 1 ...

Page 12: ...rizer on the unit using the appropriate vaporizer hardware and connect inlet outlet fittings onto the vaporizer manifold OPTIONAL ACCESSORIES ASSEMBLY INSTRUCTIONS VAPORIZER FRONT VAPORIZER REAR IMPORTANT Always refer to the user manual of the option you are installing for more detailed instructions ...

Page 13: ...equipped with a momentary valve to avoid forgetting the valve in a closed position which could cause pulmonary trauma to the animal Install the scavenger interface as per interface instructions During surgery the unused gas should be evacuated outside of the building to avoid it being inhaled by your employees and patients Connected to a gas evacuation ventilator the scavenger interface controls t...

Page 14: ...ituations It can be very useful when there is a lack of oxygen in the central system In the occasional case where you need to move an anesthetized animal to a room where there are no oxygen outlets the cylinder holder enables you to keep the patient in an ideal condition OPTIONAL ACCESSORIES ASSEMBLY INSTRUCTIONS SHELF IMPORTANT Always refer to the user manual of the option you are installing for ...

Page 15: ...hed For example if you forget the evacuation valve in a closed position the circuit alarm will alert you with a warning sound at the predetermined pressure Install wall mount adaptor as illustrated OPTIONAL ACCESSORIES ASSEMBLY INSTRUCTIONS CIRCUIT ALARM WALL MOUNT IMPORTANT Always refer to the user manual of the option you are installing for more detailed instructions ...

Page 16: ...805 0975 650 16 TM MC DISPOMED Ltd 1 800 363 1746 info dispomed com www dispomed com ASSEMBLED VIEWS 2 ...

Page 17: ...meter to approximately 200 ml Close the gas evacuation valve The unit s integrity should be verified every day and at each replacement of patient circuit components such as breathing bags tubes etc A properly maintained anesthesia machine reduces gas consumption and operating room pollution Prerequisites Oxygen supply of 40 to 60 psi 3 bar Breathing circuit and bag DAILY LEAK CHECK PROCEDURE ...

Page 18: ...eak is identified dispose of the defective item to avoid it being reused by error At the end of each day it is recommended that you close the oxygen tank and fill the anesthetic agent vaporizer DAILY LEAK CHECK PROCEDURE 5 6 7 8 Press the oxygen flush button and hold until the pressure on the manometer indicates approximately 30 cm H2 O Check the manometer The pressure must not drop But it can inc...

Page 19: ... to perform the annual maintenance Refer to this technical drawing the parts to be replaced are identified It is important to perform the daily leak check procedure page 17 once the maintenance has been completed 845 S019 x2 845 S011 x1 980 0300 180 x2 980 0300 100 x1 980 0300 105 x1 980 0300 300 x1 845 S019 x2 ANNUAL MAINTENANCE ...

Page 20: ...ONSUMABLES MATERIALS LIST 0 5 liter 940 321525 1 liter 940 321625 2 liters 940 321725 3 liters 940 321825 7g 850 0052 000 940 3704 800 5 985 0060 000 1 6 liters 905 8855 000 18 9 liters 905 8862 000 REBREATHING BAGS KRYTOX Y PATIENT CIRCUIT GAS EVACUATION TUBING CARBON DIOXIDE ABSORBENT ...

Page 21: ...y Dispomed or with the applicable codes and ordinances in place 4 Cleaned using alcohol or other damaging products Plastic rubber and consumable items are guaranteed to be free of defects at the date of delivery to the customer Any product which proves to be defective in workmanship or material will at Dispomed s discretion either be repaired or replaced Dispomed Ltd is not responsible for deterio...

Page 22: ...ed com www dispomed com ORDER PART NUMBERS Gas evacuation valve 975 0650 500 Flowmeter shield 930 0100 110 flowmeter valve 930 0105 000 Preassembled flush cartridge 975 0300 275 Dome compact 980 0300 155 Sodalime canister 980 0300 080 20 100 CM gauge 900 6050 000 ...

Page 23: ...805 0975 650 23 TM MC REORDER PART NUMBERS Fresh gas elbow 850 0030 000 Quick Connect 850 LC10004 ...

Page 24: ...rchasing a Dispomed product For further assistance contact customer support 745 Nazaire Laurin Joliette QC Canada J6E 0L6 Phone 1 800 363 1746 Canada et États Unis seulement Phone 1 450 759 9395 Fax 1 450 759 8181 info dispomed com www dispomed com Printed in Canada TM MC ...

Reviews: