22
Electromagnetic Capability
(EMC) Tables
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment
specified below. The user must assure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Electrostatic
discharge IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Guidance and Manufacturer’s Declaration -
Electromagnetic Emissions
This device is intended for use in the electromagnetic environment
specified below. The user must assure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environment
RF emissions
CISPR11
Group 1
The device uses RF energy
only for internal function.
Therefore, its RF emissions
are very low and are not likely
to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11
Class B
The device is suitable for use
in all establishments other
than domestic and those
directly connected to the
public low-voltage power
supply network that supplies
buildings for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Not
applicable