Specification Technical Description
SD2 Mode
The SD2 (Strength-Duration) mode consists of
automatic modulation intensity and pulse in a 70%
range. The intensity is always increasing while
the pulse width is decreasing and vice versa. The
intensity is decreased by 70% while the pulse
width is increased by 70% in 5 seconds. In the next
5 seconds, the intensity is increased by 70% while
the pulse width is increased by 70%. Total cycle
time is 10 seconds. Pulse rate (2-150Hz) and pulse
width (50-300 µs) are fully adjustable.
Waveform
Asymmetrical bi-phasic square pulse.
Timer
Continuous or adjustable from 1 to 60 minutes (in
1-minute increments from 1-15 and in 5-minute
increments from 15 to 60 minutes).
Record
Storage
This device can store 60 sets of operation records.
Total recorded time is 999 hours.
Voltage
0 to 50V (Load: 500 ohms)
Power
Supply
One 9V battery
Battery Life
Approximately 50 hours
Operating
Conditions
Temperature: 32°F - 104°F
Relative Humidity: 30% - 75%
Atmospheric Pressure: 525 mmHg - 795 mmHg
Size
3.98"L x 2.40"W x 0.96"H
Weight
5.29 oz. (battery included)