Direct Healthcare Services Mercury Advance SMARTcare Service Manual Download Page 19

D I R E C T H E A L T H C A R E G R O U P. C O M

S E R V I C E   M A N U A L

1 9

Contraindications For Use (Warning)

The 

Mercury Advance 

SMART

care

 System should not be used for 

patients with unstable fractures, gross oedema, burns,  

or intolerance to motion.

General Information (Caution) (Warning)

•  There are no special skills required to operate the system.

•   The Medical Professional is responsible for applying his/her 

best medical judgment when using the system.

•   The electricity supply is of the type indicated on the Control  

Unit.

•   Check the mains lead is free from damage and is positioned  

so as not to cause an obstruction, or injury. E.g. Strangulation  

of a child or trip hazard.

•   Ensure the mains lead cannot become trapped or crushed,  

e.g. by raising or lowering of the bed or bed rails or any other 

moving object.

•   The Control Unit must only be used with a suitably approved 

power cord and plug set as supplied by DHG.

•   The system is not to be used in the presence of flammable 

anaesthetics.

•  Suitable for continuous use.

•  Not suitable for sterilisation.

•   Do not position the Control Unit to make it difficult to disconnect 

the power supply or plug.

•  Do not place the System on or close to a source of heat.

•  Do not use with hot water bottles or electric blankets.

•   DHG strongly advise against smoking whilst the Control Unit is 

in use. This is to prevent accidental secondary ignition of items 

which may be flammable e.g. bed linen. The materials used in 

the manufacture of the Mercury Advance System comply with 

the required fire safety regulations.

•   Do not use sharp objects on or near the mattress system  

as this will cause damage.

•  Do not store in damp conditions.

•  Do not use in an oxygen enriched environment.

•  Not suitable for use in an Outdoor Environment.

•   Intended for both Home Healthcare and Professional Healthcare 

environments.

•  Do not connect to any other medical device or equipment.

•   Correct fuse rating MUST be used. Failure to do so could result 

in the risk of a fire.

•   The System should be cleaned after use or between patients.  

Refer to Cleaning section.

•   All internal and external hoses must be free of twists, kinks.  

The external hose should also be properly connected and 

positioned so that the risk of obstruction or injury is eliminated.

•   Do not use bleach, phenols. Chlorine based products which 

exceed 1000ppm. Solvents or alcohol based cleaners.

•   All the above warnings and cautions together with safety 

considerations should be observed at ALL times during its use.

•   Select correct setting ‘High’ or ‘Low’ as required. Care should 

be taken not to accidentally change settings once set. This may 

affect the desired requirement of the therapy. This could also be 

caused by pets, pests or children.

•    This device does not emit radiation.

13. Detachable/Removable Parts

1.  Mattress (Detached from the Control Unit by disconnecting 

the Dual Function Connector). Part No. MAT1610003 

(or variants of for the size)

2.  Electric power cable. (Removed from the Control Unit by 

pulling the cable away from the mains inlet on the side of the 

Control Unit).  

Part No. SP021016

N.B. The battery is an integral part of the PCB and is not 

removable or changeable.

Caution

Use of detachable parts not listed is not recommended by 

Direct Healthcare Group.

14. Disposal

Please refer to DHG website for recommendations and 

responsibilities for disposal within the UK WEEE guidelines.

Summary of Contents for Mercury Advance SMARTcare

Page 1: ...D I R E C T H E A L T H C A R E G R O U P C O M Service Manual...

Page 2: ...D I R E C T H E A L T H C A R E G R O U P C O M 2 M E R C U R Y A D VA N C E S M A R T C A R E...

Page 3: ...c and decontamination costs The clinical benefits of a single system are equally applicable to those of a modern hospital setting A higher maximum weight capacity up to 40 stone 254kg allows the produ...

Page 4: ...ving therapy Dependant on clinical judgement the alternating function can be operated on either a Low or High Pressure The digital Control Unit controls air flow into or out of the air cells as requir...

Page 5: ...by a trained clinician This function is used to silence the Audible Warning The LED will remain lit if the Audible Warning has been silenced previously however a fault is still detected Refer to the...

Page 6: ...located on the left of the Control Unit case Check all air hoses along the inside of the mattress each should be firmly connected Check each air cells is securely attached to its connecting pipe 3 Ch...

Page 7: ...diate 4 2 Control Unit Hang the Control Unit onto the footboard The mounting hooks swivel to suit the thickness of the footboard or rail Connect the umbilical hose to the Control Unit place the suppli...

Page 8: ...etheCPRTagtoexpelairfromthemattressquickly Note After a short period as the Mattress deflates the Low Pressure Audible Warning is activated and can be cancelled by switching the Control Unit off 5 4 T...

Page 9: ...immersetheControlUnitinwaterorotherfluids Do notautoclave norusephenolforcleaning Dowashhandsbeforecommencingthecleaningprocess Wear appropriate protective clothing such as gloves apron and a mask Ens...

Page 10: ...trol PCBA Compressor Air Outlets Rotor valve IEC Power Inlet Rotor valve motor Micro switch 1 2 3 4 5 6 7 8 6 Serial Number Identification The Control Unit is identified by serial number and GS1 compl...

Page 11: ...art by disconnecting the air tubes and the cables from the PCBA 5 Remove the rotor valve by pulling out the locking pin from the motor axes and removing the air tubes from the rotor valve 6 Remove the...

Page 12: ...lve motor Connector for Micro switch Brown blue Connector for Membrane Connector for Software download for manufacturer only Connector for red black from Power input Pressure sensor Connector for Powe...

Page 13: ...downwards and lifting the air filter holder up Take out the filter and put in a new one and put the filter holder back 2 Check the fuse Check the fuse and if required add a spare in the holder for the...

Page 14: ...nearby electronic equipment RF emission CISPR 11 Class B The system is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage p...

Page 15: ...Mains power quality should be that of a typical commercial or hospital environment Surge Immunity Test IEC 61000 4 5 1 kV line s to line s 1 kV differential mode Mains power quality should be that of...

Page 16: ...tance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur...

Page 17: ...itter m 150 KHz to 80 MHz d 1 167 P 80 MHz to 800 MHz d 1 167 P 800 MHz to 2 5 GHz d 2 333 P 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23...

Page 18: ...AINST FINGERS OR OTHER OBJECT NOT GREATER THAN 80MM IN LENGTH AND 12MM IN DIAMETER 1 PROTECTION FROM VERTICALLY DRIPPING WATER Mattress Symbols 12 Symbols Guide General Symbols REFER TO USER MANUAL SE...

Page 19: ...mattress system as this will cause damage Do not store in damp conditions Do not use in an oxygen enriched environment Not suitable for use in an Outdoor Environment Intended for both Home Healthcare...

Page 20: ...00024 Issue 2 Date February 2018 Direct Healthcare Group Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 845 459 9831 F 44 0 845 459 9832 E info directhealthcaregroup...

Reviews: