Direct Healthcare Services Mercury Advance SMARTcare Service Manual Download Page 17

D I R E C T H E A L T H C A R E G R O U P. C O M

S E R V I C E   M A N U A L

1 7

9. Technical Specification

Recommended separation distances between portable and mobile RF communications equipment and the MAT1610001  

Alternating Control Unit

The MAT1610001 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.  

The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between 

portable and mobile RF communications equipment (transmitters) and the system as recommended below, according to the maximum 

output power of the communications equipment.

Rated maximum output 

power of transmitter (W)

Separation distance according to frequency of transmitter (m)

150 KHz to 80 MHz

d = 1.167√P

80 MHz to 800 MHz

d = 1.167√P

800 MHz to 2.5 GHz

d = 2.333√P

0.01

0.117

0.117

0.233

0.1

0.369

0.369

0.738

1

1.167

1.167

2.333

10

3.689

3.689

7.379

100

11.667

11.667

23.333

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be 

estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1       At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2        These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection  

from structures, objects and people.

Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT  

or SYSTEM – for ME EQUIPMENT or ME SYSTEM that are not LIFE – SUPPORTING

Summary of Contents for Mercury Advance SMARTcare

Page 1: ...D I R E C T H E A L T H C A R E G R O U P C O M Service Manual...

Page 2: ...D I R E C T H E A L T H C A R E G R O U P C O M 2 M E R C U R Y A D VA N C E S M A R T C A R E...

Page 3: ...c and decontamination costs The clinical benefits of a single system are equally applicable to those of a modern hospital setting A higher maximum weight capacity up to 40 stone 254kg allows the produ...

Page 4: ...ving therapy Dependant on clinical judgement the alternating function can be operated on either a Low or High Pressure The digital Control Unit controls air flow into or out of the air cells as requir...

Page 5: ...by a trained clinician This function is used to silence the Audible Warning The LED will remain lit if the Audible Warning has been silenced previously however a fault is still detected Refer to the...

Page 6: ...located on the left of the Control Unit case Check all air hoses along the inside of the mattress each should be firmly connected Check each air cells is securely attached to its connecting pipe 3 Ch...

Page 7: ...diate 4 2 Control Unit Hang the Control Unit onto the footboard The mounting hooks swivel to suit the thickness of the footboard or rail Connect the umbilical hose to the Control Unit place the suppli...

Page 8: ...etheCPRTagtoexpelairfromthemattressquickly Note After a short period as the Mattress deflates the Low Pressure Audible Warning is activated and can be cancelled by switching the Control Unit off 5 4 T...

Page 9: ...immersetheControlUnitinwaterorotherfluids Do notautoclave norusephenolforcleaning Dowashhandsbeforecommencingthecleaningprocess Wear appropriate protective clothing such as gloves apron and a mask Ens...

Page 10: ...trol PCBA Compressor Air Outlets Rotor valve IEC Power Inlet Rotor valve motor Micro switch 1 2 3 4 5 6 7 8 6 Serial Number Identification The Control Unit is identified by serial number and GS1 compl...

Page 11: ...art by disconnecting the air tubes and the cables from the PCBA 5 Remove the rotor valve by pulling out the locking pin from the motor axes and removing the air tubes from the rotor valve 6 Remove the...

Page 12: ...lve motor Connector for Micro switch Brown blue Connector for Membrane Connector for Software download for manufacturer only Connector for red black from Power input Pressure sensor Connector for Powe...

Page 13: ...downwards and lifting the air filter holder up Take out the filter and put in a new one and put the filter holder back 2 Check the fuse Check the fuse and if required add a spare in the holder for the...

Page 14: ...nearby electronic equipment RF emission CISPR 11 Class B The system is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage p...

Page 15: ...Mains power quality should be that of a typical commercial or hospital environment Surge Immunity Test IEC 61000 4 5 1 kV line s to line s 1 kV differential mode Mains power quality should be that of...

Page 16: ...tance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur...

Page 17: ...itter m 150 KHz to 80 MHz d 1 167 P 80 MHz to 800 MHz d 1 167 P 800 MHz to 2 5 GHz d 2 333 P 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23...

Page 18: ...AINST FINGERS OR OTHER OBJECT NOT GREATER THAN 80MM IN LENGTH AND 12MM IN DIAMETER 1 PROTECTION FROM VERTICALLY DRIPPING WATER Mattress Symbols 12 Symbols Guide General Symbols REFER TO USER MANUAL SE...

Page 19: ...mattress system as this will cause damage Do not store in damp conditions Do not use in an oxygen enriched environment Not suitable for use in an Outdoor Environment Intended for both Home Healthcare...

Page 20: ...00024 Issue 2 Date February 2018 Direct Healthcare Group Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 845 459 9831 F 44 0 845 459 9832 E info directhealthcaregroup...

Reviews: