PDA-100 Photodiode Array Detector
2-10
Doc. 031644-02 9/03
2.5
Rear Panel
Figure 2-5 illustrates the rear panel of the PDA-100.
Power Entry
The power entry is socketed for a modular power cord (IEC 320 C13). The PDA-
100 operates from input voltages over a range from 85 Vac to 264 Vac, 47 Hz to
63 Hz power and requires a grounded, single-phase power source. The input
power is 162 W, maximum and the output power is 130 W, maximum. The
appropriate line voltage and frequency are selected automatically.
Figure 2-5. PDA-100 Rear Panel
SHOCK HAZARD—To avoid electrical shock, use a grounded
receptacle. Do not operate the PDA-100 or connect it to AC power
mains without an earthed ground connection.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the PDA-100 and is easily
accessible.
Operation at AC input levels outside of the specified operating voltage
range may damage the PDA-100.
(+)
OUT
TTL
RELAY OUT
2
1
4
3
2
1
2
1
2
1
12
11
10
9
8
7
6
5
4
3
2
1
DX-LAN
ANALOG OUTPUT
CHANNEL 4
CHANNEL 3
CHANNEL 2
CHANNEL 1
TTL IN (+)
TTL
GND
(+)
056338_01
FOR CONTINUED PRO-
TECTION AGAINST
RISK OF FIRE OR
SHOCK, REPLACE
ONLY WITH SAME
TYPE AND RATING
OF FUSE.
WARNING
DUAL FUSE INPUTS.
REMOVE POWER CORD
BEFORE SERVICING.
H3.15A
250V
XXXXXXXX
TUV
MODEL: PDA-100
INPUT: 100-240V
50/60Hz, 1.5A
SERIAL NUMBER
035779-05
Made in U.S.A.
Dionex Corporation, Sunnyvale, California 94086
An alternative license is available, on reasonable terms, to practice any of the foregoing
methods of use which can be praticed without infringing patented apparatus or products.
Apparatus and products manufactured and sold by Dionex Corporation under one or more
of the following U.S. Patents owned or licensed by Dionex: 4,242,097; 4,252,644; 4,314,823;
4,351,909; 4,383,047; 4,455,233; 4,474,664; 4,482,967; 4,512,245; 4,644,807; 4,700,059; 4,751,004;
4,839,788; 4,939,924; 4,999,098; 5,248,426; 5,271,903; 5,296,115; 5,324,752; 5,358,612; 5,457,626;
5,647,967; 5,660,727, and all equivalents in other countries where issued. Purchaser is granted
a paid-up, non-exclusive license to practice with, and for the useful life of, this particular
apparatus or product, the methods of use in any of the foregoing patents and in U.S. Patents
5,126,272; 4,265,634; 4,414,842 and all equivalents in other countries where issued.
Fuse Holder
Main Power
Receptacle
PDA-100
Serial Number
Relay/TTL
Connectors
DX-LAN
Connector
Analog Outputs
Drain
Tube
LAN Identification
Number
057355-01
LAN ID
PDA-100
IC
P/N 057060