Due to the possible occurrence of an electrostatic charge on the non-conductive parts of 

the sensor, it is necessary to ground all sensors intended for use in environments with 

an explosion hazard DLM–35Xi (XiT, XiM, XiMT). This can be performed by grounding el. 

conductive tanks or el. conductive tank lids, and in the case of el. non-conductive tanks 

using and grounding an auxiliary plate electrode PDE-27.
In the event that the level meter (sensor) is installed in an outdoor environment at a 

distance greater than 20 m from the outdoor switchboard, or from an enclosed building, 

it is necessary to supplement the electrical cable leading to the level meter (sensor) with 

suitable overvoltage protection.
In the event of strong ambient electromagnetic interference, paralleling of conductors with 

power distribution, or for distribution to distances over 30 m, we recommend using a shielded 

cable and grounding the shielding on the side of the power source.

7 .  Control elements

Cable gland

or connector

Orange LED

Sensitive spot 

+

Sensitive spot 

-

Green LED

spot 

 

•  input to setting mode
•  direct setting to value 4 mA (0 V)
•  decrease in values along defined steps

spot 

•  input to setting mode
•  direct setting to value 20 mA (10 V)
•  increase in values along defined steps

Fig. 7: Top view of level sensor

8 .  Function and status indication

LED indicator colour

function

"RUN"

green

Measuring function indicator

flashing

 – (repeats according to period of measuring approx. 0.5 s) 

– correct function of measuring the level

dark

–  incorrect installation or malfunction. 

  LED is also not illuminated in limit setting mode.

alternating flashing green and orange LED

 – incorrectly set limits

"STATE"

orange

Settings indication

slow flashing

 – indication of setting the lower limit

rapid flashing

 – indication of setting the upper limit

permanent shine

 –  the level meter is prepared to confirm the limit 

setting by means of mag. pen 

3x short flashes 

– setting confirmation

simultaneous shine of green and orange LED

 – during touching 

the mag. pen, when the limit setting is confirmed

Summary of Contents for DLM-35

Page 1: ...ed in this manual before the first use of the level meter Keep the manual at a safe place The manufacturer reserves the right to implement changes without prior notice pr myslov elektronika Capacitive...

Page 2: ...Control elements 11 8 Function and status indication 11 9 Settings 12 10 Order code 14 11 Correct specification examples 14 12 Accessories 15 13 Safety protections compatibility and explosion proof 15...

Page 3: ...iT XiMT available DLM are offered in variants with various types of process con nection metric and pipe thread pressure thread NPT 1 Basic description To ensure maximum safety of control processes we...

Page 4: ...30 Uncoated rod electrode used for level measurement of bulk solids cement flour sand plastic granulate and electrically non conductive liquids vegetable oil diesel fuel petrol Maximum electrode lengt...

Page 5: ...PT 3 4 All dimensions are in mm 64 22 35 16 36 6 5 E100 E1000 10 10 8 15 42 DLM 35_ 20 DLM 35_ 21 22 25 DLM 35_ 30 DLM 35_ 31 thread E100 E1000 10 10 6 46 46 E100 E3000 E100 E3000 8 Measuring electrod...

Page 6: ...r M12 6 5 12 18 12 20 15 20 M12 13 Variant H with cable gland for protected hoses for using in an outdoor area or in area with increased moisture Variant V with plastic cable gland with spiral relief...

Page 7: ...with steel armouring For level measurement of material in plastic and glass vessels by level meter without reference tube is necessary to connect grounding screw at housing with auxiliary electrode wh...

Page 8: ...th reference tube DEEP METAL AND CONCRETE HOPPERS For types DLM 35_ 50 connection to the armouring E electrode length mm select so that the end of the electrode would be at least 20 mm under the lowes...

Page 9: ...cable for the Xi XiT XiM XiMT variations must be selected with respect to the maximum permitted parameters usually inductance and capacity of the outside intrinsically safe circuit of supply units IRU...

Page 10: ...he shielding on the side of the power source 7 Control elements Cable gland or connector Orange LED Sensitive spot Sensitive spot Green LED spot input to setting mode direct setting to value 4mA 0V de...

Page 11: ...dicators STATE and RUN it is possible to put away the magnetic pen The LED indicator STATE will briefly flash 3x to confirm the setting of the limit Setting the upper limit 20 mA 10 V Direct 1 Fill th...

Page 12: ...irect 1 Bring the level in the tank to a known determinable level best when close to the maximum measured level 2 Touch the magnetic pen for approx 3 seconds on the sensitive spot of the level meter A...

Page 13: ...ectrode length 200mm process connection DLM 35 electrode length in mm E type of output I current 4 20 mA U voltage 0 10 V connection method A short stainless steel gland cable length B plastic threade...

Page 14: ...es for safety and electromagnetic compatibility Special conditions for safe use of variant DLM 35Xi XiT XiM XiMT Level meters DLM 35Xi XiT XiM XiMT are intended for connection to approved spark safe p...

Page 15: ...g operation and maintenance of the DLM 35 level meter has to be carried out in accordance with this instruction manual the provisions of regulations in force regarding the installation of electrical e...

Page 16: ...T 2 4 Ga Gb II 1 D Ex ia II T C 2 I C 120 Da Db FTZ 16 ATEX 0138X No ______ E_____ Cable __m II 1 G Ex ia IIB T4 Ga II 1 D Ex ia II T C I C 120 Da FTZ 16 ATEX 0138X 0 V U 1 bn 3 bu 0 V U 1 bn 3 bu 0...

Page 17: ...t 0 5 C a 4 7 1026 760 01 Zl n Czech Republic w t tek pro varianty DLM 35XiM __ __ I _ __ t tek pro varianty DLM 35XiMT __ __ I _ __ No ______ DLM 35XiM __ __ I _ __ E_____ Cable __m No ______ DLM 35X...

Page 18: ...dielectric strength 47 nF 350 V AC Ambient temperature range 40 85 C Protection type DLM 35_ _ _ C _ _ type DLM 35_ _ _ A B V H _ _ IP67 IP68 Maximum load resistance for current output at U 24 V Rmax...

Page 19: ...nits DLM 35XiMT Intrinsically safe high temperature explosion proof performance for use in mines with the occurrence of methane or coal dust I M1 Ex ia I Ma with intrinsically safe supply units Use ma...

Page 20: ...50 40 C 250 C 40 C 85 C 40 C 85 C DLM 35NT 20 30 40 C 300 C 40 C 200 C 40 C 85 C DLM 35NT 21 22 31 40 41 40 C 200 C 40 C 200 C 40 C 85 C DLM 35NT 25 40 C 200 C 40 C 200 C 40 C 85 C DLM 35NT 50 40 C 2...

Page 21: ...LM 35NT 25 2 0 MPa 20 bar 2 0 MPa 20 bar 2 0 MPa 20 bar 2 0 MPa 20 bar 2 0 MPa 20 bar DLM 35NT 50 0 1 MPa 1 bar 0 1 MPa 1 bar 0 1 MPa 1 bar 0 1 MPa 1 bar 0 1 MPa 1 bar DLM 35Xi XiM 20 30 5 MPa 50 bar...

Page 22: ...r to resolve the situation If the device is to be further shipped it must be wrapped in its original packaging and protected against impact and weather conditions Store the device in its original pack...

Page 23: ...r o U Tescomy 249 760 01 Zl n Czech Republic Phone 420 577 002 002 Fax 420 577 002 007 E mail sale dinel cz www dinel cz Find the updated version at www dinel cz version 08 2017 industrial electronics...

Reviews: