Key Functions of Balance 

ON/OFF      switch key 
UNIT            unit conversion key 
PCS              counting key 
CAL              calibration key 
TAR              taring & zere setting key 

 

Installation of Balance 

 

Place the balance on a  flat  workbench, shaking of balance, flow  and 
strong electromagnetic interference should be avoided.   

 

III Turning On

 

Plug the adapter or install 4 AA batteries(4×1.5vAA battery).Press the 

ON/OFF  key  to  display  “888888”.  After  “300.000”displayed,  it 

enterinto  60  seconds  count  down  of  preheating.  At  last  it  displays 

0.000g.  Occurs  when  the  upper-right  corner  zero  “O”  symbol,  it 

means zero has stabilized. 

 

 

-1- 

. Calibration 

 

1.

 

Calibration  should  be  taken  after  turning  on  and  preheating  for 

10-30 minutes. 

2.

 

Calibration: 

In  the  case  of  the  balance  doesn’t  load  any  object,  press  “CAL” 

calibration, after 5 seconds it display “—CAL—”, release the key, 

later  it  displays  “calibration  weight  value”  is  flashing.  Now  put 

the weight that value same with “calibration weight value” on the 

pan.  After  second  minutes  it  displays  “calibration  weight  value”, 

remove the weight,  the  balance  return to  zero  automatically after 

dilpays  “------”  again.  Calibration is  complete.  If still slight  error 

after  first  calibration,  please  re-calibrate  according  to  aboved 

process several times or carry out linear calibration. 

3.

 

Liner Calibration 

In  the  case  of  the  balance  doesn’t  load  any  object,  press  “CAL” 

calibration, after 5 seconds it display “—CAL—”, release the key, 

press  “CAL”key  and  don’t  release,    corresponding  linear 

calibration  values  for  each  sections  will  display  after 

approximately  5  seconds.  The  flashing  value  means  value  of 

weights  shall  be  loaded,  upon  loading  the  displayed  value  is  not 

blinking, it means that the balance has memorized segment linear 

nominal value. Reomove the weights and wait for next period of     

                                                 

-2-

                     

                                                 

 