DigiVac 2L760 Instruction Manual Download Page 8

 

   DIGIVAC 2L760 Instruction Manual Page 8 

 

6.0    SERVICING -

 GAUGE TUBE CLEANING. 

            
In some cases, a gauge tube may become contaminated with oil or other foreign matter. Once contamination 
occurs, the sensor needs to be replaced.  Please send the controller back to DIGIVAC for mating gauge tube 
replacement. 
 
 
6.1   FACTORY REPAIR AND CALIBRATION. 
 
The vacuum gauge assembly is designed to provide years of trouble-free service, and the liberal internal use of 
plug-in components makes it easily repairable.  Field servicing of the unit is not recommended. The accuracy of 
calibration is dependent on the quality of the gauge tube, accuracy of calibration is difficult to achieve in a field 
setting. Each DIGIVAC vacuum gauge controller is calibrated to the particular vacuum gauge sensor that is 
shipped with the unit.   
 
In the event the user must recalibrate the instrument, the following procedures are suggested. 

A.

 

Before re-calibrating the instrument, it should be ascertained that the instrument is in fact incorrect. In 
many cases, the problem will be with a tube that is contaminateded, or a system that is operating 
improperly. It is recommended that a spare tube be kept on hand and stored in a clean, dry place. Then, 
in cases of suspect readings, the tube should be changed before proceeding further. 

 

B.

 

Calibration instructions: 
1.

 

Put gauge on a manifold with a standard such as a capacitance manometer, and set manifold 
pressure to 750 Torr or 1000 mbar 

2.

 

Adjust “ATM” potentiometer to 750 Torr (for Torr units) (or 1000 mbar) 

(matching standard) 

3.

 

Set manifold to 5 Torr 

and adjust “vac” potentiometer so it reads 5 Torr +/

- 0.2 Torr.  (for mbar 

installations, set at 6.4 mbar +/- 0.2 mbar) 

 
7.0   NOTES ON CALIBRATION. 

 
The DIGIVAC is calibrated in nitrogen.  This device can measure any clean dry gas and retain calibration. 
 
Note that isolated transducers like capacitance manometers are independent of gas type or moisture content. 
 

8.0   UNDERSTANDING TORR. 

 
The 2L760 and many similar instruments are calibrated in microns or "millitorr." Microns are not really a measure 
of vacuum at all, but rather of absolute pressure.  It will be recalled that the pressure of the atmosphere is 
14.696 or approximately 14.7 pounds per square inch at sea level. This pressure is due to the weight of all of the 
air in the earth's atmosphere above any particular square inch.  This 14.696 psi is equivalent to the pressure 
produced by a mercury column of approximately 29.92 inches high or .76 meters (about 3/4 of a yard) or 760 
millimeters of mercury.  Atmospheric pressure varies greatly with altitude. It decreases approximately 1 inch of 
mercury per thousand feet of altitude. It also varies widely with local weather conditions. (Variations of one half 
inch in a single day are common.)  The word vacuum means pressure lower than atmospheric or "suction," but, 
in describing negative pressure, the atmosphere is only a satisfactory reference if we are dealing with values of 
vacuum down to about 27 inches of mercury. Below that, it is much more useful to talk in terms of absolute 
pressure, starting from absolute zero. The 2L760 and all similar instruments measure negative pressure below 

27” of mercury.

 

 
One TORR, a commonly used unit, is an absolute pressure of one millimeter of mercury. A millitorr is equal to 
one thousandth of a TORR. A MICRON is the same as a millitorr.  

Summary of Contents for 2L760

Page 1: ...continued However we are able to service calibrate and or repair existing Models in the market at this time If you require a replacement gauge we recommend the StrataVac 775i see web page here that ha...

Page 2: ...nd Principle of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolut...

Page 3: ...m against a NIST standard 2 0 CONSTRUCTION The controller consists of the indicating and controlling instrument the configured sensor the sensor cable the output interfaces and an univeral AC power ad...

Page 4: ...tions The RS232 connection can be made to a PLC or computer via a male DB9 cable connection to the female jack on the DIGIVAC The DIGIVAC acts as a DCE so a straight serial connection is appropriate P...

Page 5: ...otentiometer to the desired vacuum level 3 Follow the same procedure for setting set point 2 shorting the center pin and the right pin then adjusting the potentiometer on the right 4 Replace the jumpe...

Page 6: ...anual Page 6 Feel free to connect to the IP address 10001 with HyperTerminal putty or your favorite communication tool to communicate directly with the device To use the 2L760 web terminal capability...

Page 7: ...the usercan execute all the commands in Press return or enter to submit the query or change the set point Note that while the web terminal is open the main 2L760 screen will not update If at any time...

Page 8: ...adjust vac potentiometer so it reads 5 Torr 0 2 Torr for mbar installations set at 6 4 mbar 0 2 mbar 7 0 NOTES ON CALIBRATION The DIGIVAC is calibrated in nitrogen This device can measure any clean d...

Page 9: ...losure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC Ins...

Page 10: ...LAR PURPOSE NON INFRINGEMENT OR ANY OTHER WARRANTY EXPRESSED OR IMPLIED LIMITATION OF LIABILITY You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect puni...

Reviews: