DigiVac 2L760 Instruction Manual Download Page 4

 

   DIGIVAC 2L760 Instruction Manual Page 4 

The instrument should be located in a clean, dry environment for best results.   The gauge tube cable should be 
identified by wire tags or markings specific to your environment.

 

 

Sensors perform best when installed in a thread-down orientation in a clean, dry vacuum system.  While 
threading the gauge tube into the manifold, the gauge tube cable should be disconnected to avoid damage.  In 
this way, twisting of the cable and the octal socket on the tube is avoided.  Care should be exercised to install the 
tubes in a dry part of the system. The gauge tube should be protected against oil and other contaminants by 
installing it in such a way to protect it.  A good practice is to mount the gauge tube in the most vertically distant 
place from oil and other contaminants as applicable.  The gauge tube should be mounted in the most stable 
pressure region of the vessel to be measured.  For example, it would be better to install the gauge tube on a tank 
rather than on the pipe that is directly connected to a vacuum pump.  In the event of contamination, see section 
6.0 for gauge tube cleaning instructions. 
 
The RS232 connection can be made to a PLC or computer via a male DB9 cable connection to the female jack on 
the DIGIVAC.  The DIGIVAC acts as a DCE, so a straight serial connection is appropriate. 
 
Please use the supplied 5V AC adapter with your instrument.  This adapter provides clean short protected power 
to protect and insure accuracy of the internal circuitry. 

 
 
5.0   OPERATION. 

 
After installation, the DIGIVAC is ready for operation.  The unit will normally provide accurate readings 
immediately.   
 
Make all connections to sensors, and relay outputs with the power disconnected.  NEVER DISCONNECT SENSOR 
OR OUTPUT WIRES WHILE UNIT IS POWERED UP. 
 
In cases where the system has contaminants, as is often the case with metalizing and coating equipment, it is 
often effective to isolate the gauge tube with a solenoid or manual valve during periods when contamination is 
most active. 

 

 

 

 
The user can purchase additional options: 
 

 

RS232 

 The instrument puts out a standard RS232 serial stream with settings 9600, 8, N, and 1.  The 

unit transmits but does not receive, and displays the current vacuum indication in the current units. 

 

Analog out 

 This output reads from 0 to 5 Volts from a pressure of 1 Torr to 760  Torr.   

 

Relay Control outputs 

 
Units configured with controls will have two set point SPDT relays adjustable with a potentiometer.   
 

 
 
 
Operation: 

Summary of Contents for 2L760

Page 1: ...continued However we are able to service calibrate and or repair existing Models in the market at this time If you require a replacement gauge we recommend the StrataVac 775i see web page here that ha...

Page 2: ...nd Principle of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolut...

Page 3: ...m against a NIST standard 2 0 CONSTRUCTION The controller consists of the indicating and controlling instrument the configured sensor the sensor cable the output interfaces and an univeral AC power ad...

Page 4: ...tions The RS232 connection can be made to a PLC or computer via a male DB9 cable connection to the female jack on the DIGIVAC The DIGIVAC acts as a DCE so a straight serial connection is appropriate P...

Page 5: ...otentiometer to the desired vacuum level 3 Follow the same procedure for setting set point 2 shorting the center pin and the right pin then adjusting the potentiometer on the right 4 Replace the jumpe...

Page 6: ...anual Page 6 Feel free to connect to the IP address 10001 with HyperTerminal putty or your favorite communication tool to communicate directly with the device To use the 2L760 web terminal capability...

Page 7: ...the usercan execute all the commands in Press return or enter to submit the query or change the set point Note that while the web terminal is open the main 2L760 screen will not update If at any time...

Page 8: ...adjust vac potentiometer so it reads 5 Torr 0 2 Torr for mbar installations set at 6 4 mbar 0 2 mbar 7 0 NOTES ON CALIBRATION The DIGIVAC is calibrated in nitrogen This device can measure any clean d...

Page 9: ...losure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC Ins...