DigiVac 200 Pirani Instruction Manual Download Page 7

 

   DIGIVAC 201 series Instruction Manual Page 7 

 
 

6.0    SERVICING -

 GAUGE TUBE CLEANING. 

            
In many cases, a gauge tube may become fouled with oil or other foreign matter. It is often possible to 
restore the functionality of contaminated probes with cleaning. If the contaminant is known, the tube 
should be filled with a fluid that is known to be a solvent to that contaminant. As an example, ether is 
often effective in removing residues of some oils.  Commercial carburetor cleaners are very powerful 
solvents and are highly effective against some contaminants. 
 
After cleaning with solvents, the gauge tube should be completely dried or flushed with a volatile solvent 
to assure that it is dry prior to re-installing it. If this is not done, contamination of the system may result. 
 
 
6.1   FACTORY REPAIR AND CALIBRATION. 
 
The vacuum gauge assembly is designed to provide years of trouble-free service, and the liberal internal 
use of plug-in components make it easily repairable.  No field servicing of the unit is recommended, 
other than replacement of the gauge tube, but factory servicing and calibration are available at a nominal 
cost and turn-around times of 24 hours are typical.   
 
6.2   FIELD CALIBRATION. 
 
Each DIGIVAC vacuum gauge controller is calibrated to the particular vacuum gauge sensor that is 
shipped with the unit.  While changing the gauge tube is possible, it will result in a slightly different 
reading as all gauge tubes are not created equal.  Calibration for this unit must be performed by  
DigiVac. 
 
Before re-calibrating the instrument, it should be ascertained that the instrument is in fact incorrect. In 
many cases, the problem will be with a tube that is fouled, or a system that is operating improperly. It is 
recommended that a spare tube be kept on hand and stored in a clean, dry place. Then, in cases of 
suspect readings, the tube should be changed before proceeding further. 

 
7.0   NOTES ON CALIBRATION. 

 
The DIGIVAC is calibrated in nitrogen, which has thermal properties virtually identical to air. If you are 
using a Pirani vacuum gauge, other gases will affect the readings by an amount proportional to the 
thermal conductivity of the gases. In most cases, the gases present in a vacuum system will be air, 
nitrogen, or oxygen, and no appreciable errors will occur. 
 
Certain other gases, however, have thermal conductivity significantly greater than air and will cause the 
instrument to read higher than the actual amount of pressure. Examples of such gases are water vapor, 
fluorocarbon refrigerants, and acetone. Conversely, other gases have thermal conductivity significantly 
lower than air and will cause the instrument to read lower than actual pressure. Examples of such gases 
include helium, oxygen and to a lesser extent, CO2. 
 
When interpreting readings using gases other than air, it should be borne in mind that the DIGIVAC 
reads Millitorr, which is a measure of absolute pressure - that is the opposite of vacuum. Thus, a lower 
numerical reading actually is a higher level of vacuum. For more information, refer to section 8.0. 
 
Note that isolated transducers like capacitance manometers are independent of gas type. 
 

8.0   UNDERSTANDING TORR. 

 
The DIGIVAC and many similar instruments are calibrated in microns or "millitorr." It is appropriate to 
discuss what microns are and to relate microns to other measures of pressure and vacuum. 

Summary of Contents for 200 Pirani

Page 1: ...AC Model 200 Pirani Digital Vacuum Gauge Revsion K13A YOU MUST READ THIS MANUAL BEFORE USE Readings from 1 2000 millitorr The DIGIVAC Company 105B Church Street Matawan NJ 07747 www DIGIVAC com 732 76...

Page 2: ...le of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolute Pressure...

Page 3: ...rument the sensor configured the sensor cable the output interfaces and an international AC power adapter The instrument is housed in a rugged free standing vinyl clad metal enclosure It can be placed...

Page 4: ...V AC adapter with your Instrument This adapter provides clean short protected power to protect and insure accuracy of the internal circuitry 5 0 OPERATION After installation the DIGIVAC is ready for i...

Page 5: ...200 Surf over to http 192 168 0 200 or whatever your IP address was set to and say yes to security screens authorizing a DIGIVAC applet to run in your browser You should see a screen that looks like...

Page 6: ...in Press carriage return or enter to submit the query or change the set point Note that while the web terminal is open the main 200 Pirani screen will not update If at any time the web terminal or th...

Page 7: ...at the instrument is in fact incorrect In many cases the problem will be with a tube that is fouled or a system that is operating improperly It is recommended that a spare tube be kept on hand and sto...

Page 8: ...altitude It also varies widely with local weather conditions Variations of one half inch in a single day are common The word vacuum means pressure lower than atmospheric or suction but in describing...

Page 9: ...closure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC I...

Page 10: ...CULAR PURPOSE NON INFRINGEMENT OR ANY OTHER WARRANTY EXPRESSED OR IMPLIED LIMITATION OF LIABILITY You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect pu...

Reviews: