1.0   DESCRIPTION AND PRINCIPLE OF OPERATION. 

 
The Model 200 Pirani is a full featured vacuum instrument that is fast, accurate and rugged. This unit 
builds on the engineering philosophy that made the Model 200 our best selling vacuum instrument: 
simple is best and robust is a must. The model 200 Pirani employs a VRC Pirani tube instead of a 
Varian 531 gauge tube.  The instrument features large, bright LEDs visible from across the room with a 
hard, vinyl clad metal enclosure. This entry into our gauge line-up, enables the user to choose yet 
another sensor option to best meet their vacuum requirements. The straight forward design yields a low 
power instrument, that is simple to build and simple to operate. The low power feature reduces thermal 
effects, to improve repeatability. Unlike many other Pirani gauges on the market, this unit has been 
recently redesigned with newer, more reliable components to insure a long product life. The Model 200 
Pirani can be ordered with optional control relays, analog output, and RS232 output, to simplify 
integration with existing process controllers or computers. In addition, this device can be Ethernet 
enabled, to allow monitoring via a web browser with our free ReAct browser software. Get access to your 
instrument from across the hall, or across the world. All instruments leave our factory tested and 
calibrated under real vacuum against a NIST standard. 
         

2.0   CONSTRUCTION. 

 
The controller consists of the indicating and controlling instrument, the sensor configured, the sensor 
cable, the output interfaces and an international AC power adapter.   
 
The instrument is housed in a rugged free-standing vinyl clad metal enclosure. It can be placed on a 
suitable surface for easy viewing from across the room.  The gauge tube houses the various Pirani 
sensing, heating and compensating elements and terminates in an octal connector. On this model, the 
connector wiring terminates at the instrument with a 9 pin DSub connection.  Regulating circuitry in the 
DIGIVAC provides constant current for gauge tube excitation, and thus compensates for resistance in 
the probe leads.  
 

3.0   UNPACKING AND INSPECTION. 

 
After the instrument is received, it should be carefully unpacked and inspected for damage during 
shipment and for completeness. The package should contain, as a minimum, the instrument, the sensor 
and cable ordered, an AC power adapter and an instruction manual. In the event of a loss during 
shipment, a claim should immediately be made to the common carrier or the postal service, as 
applicable. The DIGIVAC warranty pertains only to the instrument, and does not cover losses in 
shipping.  Each 200 Pirani should come with: 

o

 

User Manual (this document) 

o

 

Display controller (gray box with red lens)

 

o

 

Power supply (international adapter with US plug) 

o

 

Configured Sensors 

o

 

RS232 (optional) 

o

 

RS232 Port 

o

 

5

’ RS232 Cable (phono to Female DB9) 

o

 

Ethernet (optional, includes RS232) 

o

 

Ethernet Port 

o

 

Null modem RS232 converter 

o

 

2’ Ethernet Cable 

o

 

Null Modem Ethernet Cable 

o

 

Configuration CD 

 
 

4.0  INSTALLATION. 

The instrument should be located in a clean, dry environment for best results.   The gauge tube cable 
should be identified by wire tags or markings specific to your environment.

 

 

Summary of Contents for 200 Pirani

Page 1: ...AC Model 200 Pirani Digital Vacuum Gauge Revsion K13A YOU MUST READ THIS MANUAL BEFORE USE Readings from 1 2000 millitorr The DIGIVAC Company 105B Church Street Matawan NJ 07747 www DIGIVAC com 732 76...

Page 2: ...le of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolute Pressure...

Page 3: ...rument the sensor configured the sensor cable the output interfaces and an international AC power adapter The instrument is housed in a rugged free standing vinyl clad metal enclosure It can be placed...

Page 4: ...V AC adapter with your Instrument This adapter provides clean short protected power to protect and insure accuracy of the internal circuitry 5 0 OPERATION After installation the DIGIVAC is ready for i...

Page 5: ...200 Surf over to http 192 168 0 200 or whatever your IP address was set to and say yes to security screens authorizing a DIGIVAC applet to run in your browser You should see a screen that looks like...

Page 6: ...in Press carriage return or enter to submit the query or change the set point Note that while the web terminal is open the main 200 Pirani screen will not update If at any time the web terminal or th...

Page 7: ...at the instrument is in fact incorrect In many cases the problem will be with a tube that is fouled or a system that is operating improperly It is recommended that a spare tube be kept on hand and sto...

Page 8: ...altitude It also varies widely with local weather conditions Variations of one half inch in a single day are common The word vacuum means pressure lower than atmospheric or suction but in describing...

Page 9: ...closure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC I...

Page 10: ...CULAR PURPOSE NON INFRINGEMENT OR ANY OTHER WARRANTY EXPRESSED OR IMPLIED LIMITATION OF LIABILITY You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect pu...

Reviews: