DigiVac 200 Pirani Instruction Manual Download Page 1

 
 
 
 
 
 
 
 
 

 

 
 
 

INSTRUCTION MANUAL 

 

DIGIVAC Model 200 Pirani 

 

Digital Vacuum Gauge  

Revsion K13A

 

 

YOU MUST READ THIS MANUAL BEFORE USE 

 

Readings from 1 

– 2000 millitorr

 

  
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 

 

 

 

 

 
 
 
 
 

 

 

 

 

 

 

 

The DIGIVAC Company  

 

 

 

 

 

 

 

 

 

105B Church Street 

 

 

 

 

 

 

 

 

 

Matawan NJ  07747 

 

 

 

 

 

 

 

 

www.DIGIVAC.com 

 

 

 

 

 

 

 

 

(732) 765-0900 

 

 

 

 

 

 

 

 

(732) 765-1800 FAX 

 
 
 

 

 

 

 

 

 

 

 

 

 

 
 

Summary of Contents for 200 Pirani

Page 1: ...AC Model 200 Pirani Digital Vacuum Gauge Revsion K13A YOU MUST READ THIS MANUAL BEFORE USE Readings from 1 2000 millitorr The DIGIVAC Company 105B Church Street Matawan NJ 07747 www DIGIVAC com 732 76...

Page 2: ...le of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolute Pressure...

Page 3: ...rument the sensor configured the sensor cable the output interfaces and an international AC power adapter The instrument is housed in a rugged free standing vinyl clad metal enclosure It can be placed...

Page 4: ...V AC adapter with your Instrument This adapter provides clean short protected power to protect and insure accuracy of the internal circuitry 5 0 OPERATION After installation the DIGIVAC is ready for i...

Page 5: ...200 Surf over to http 192 168 0 200 or whatever your IP address was set to and say yes to security screens authorizing a DIGIVAC applet to run in your browser You should see a screen that looks like...

Page 6: ...in Press carriage return or enter to submit the query or change the set point Note that while the web terminal is open the main 200 Pirani screen will not update If at any time the web terminal or th...

Page 7: ...at the instrument is in fact incorrect In many cases the problem will be with a tube that is fouled or a system that is operating improperly It is recommended that a spare tube be kept on hand and sto...

Page 8: ...altitude It also varies widely with local weather conditions Variations of one half inch in a single day are common The word vacuum means pressure lower than atmospheric or suction but in describing...

Page 9: ...closure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC I...

Page 10: ...CULAR PURPOSE NON INFRINGEMENT OR ANY OTHER WARRANTY EXPRESSED OR IMPLIED LIMITATION OF LIABILITY You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect pu...