Main Amplifier Front Panel & Headstage Printed Symbols
Main amplifier power indicator
+
Reference electrode connection
Caution, consult accompanying
documents.
-
Active electrode connection
Signal input connection.
COM
Common electrode connection
Connection to D360R-4
headstage
Type BF applied part
Consult operator documentation
(this manual) before use of this
equipment.
Headstage power indicator
Rear Panel Components
1.
Power On/Off Switch
Two position rocker switch
2.
Mains Voltage Selector
Selectable between 115V and 230V. Ensure appropriate mains fuses are fitted before powering on.
3.
Potential Equalisation Connector (PEC)
Earth/Ground reference for unit and bonding point. This is to be used when the earth/ground
conductor in the mains cord of all units in use cannot be relied upon or is not available.
8
7
6
1
5
4
3
2