Description of display 
 

LED 

Name 

Description 

BAT 

Low battery sign  Battery Voltage is low 

STB 

Stable sign 

Display show weight is stable 

GROSS 

Gross weight 

Display the gross weight 

NET 

Net weight 

Display value show net weight after deduct the 
container from the platter 

TARE 

Tare weight 

Tare weight is after deduct container weight 

ZERO 

Zero 

Display weight is zero 

Kg 

Unit kg 

Displayed weight is in Kilogram (kg) 

LB 

Unit lb 

Displayed weight is in Pounds (lb) 

Oz 

Unit oz 

Displayed weight is in Ounce (oz) 

Ct 

Unit ct 

Displayed weight is in carat (ct) 

Negative 

Negative weight 

Displayed weight is (-) negative 

Charge 

Charged battery 

1.  When turn to Red LED show battery is LOW 
2.  When turn to Green LED show battery is full 

HI 

Over limit 

The weight is over the preset value 

OK 

Pass limit 

The weight I between HI & LO preset value 

LQ 

Low limit 

The weight is under the preset value 

 

Switch bottom function : 
 

1.  Switches Power On / Off. 

i.

 

Press 

I

 once, the display show a self-test and come to Zero on the screen. 

Each LED will display 0 to 9 in sequence during the self-test. 

ii.

 

Press and hold  

I

   for 3.0 second, the display will shut down automatically. 

2. 

Place the container on the platter .Press      

T   

the weight of the container will be 

subtracted and “ NET “ & “ 0 “ will appear . 

3. 

Re-zero the weight and display show zero. 

4. 

Press 

to choose the unit weight and display will show: Kg, lb, oz and ct. 

5.  Press 

→ 

the display shown number of times weighed in “ n x x ” and after to 

print out accumulated total. 

i. 

To  clear  or  reset  the  accumulated  total  ,  press 

once.  Press 

↑    to  print 

current weight and the weight will be added to the accumulated total. 

ii. 

To clear or reset the accumulated totals, press 

Summary of Contents for DI-28SS

Page 1: ...iko Co Ltd Head office 13 12 Kugahara 5 chome Ohta Ku Tokyo 146 8580 Japan TEL 81 3 3752 2131 FAX 81 3 3752 2801 Ohsaki office 2 3 13 Ohsaki Shinagawa Ku Tokyo 141 8566 Japan TEL 81 3 5496 7600 FAX 81 3 5496 7620 http www digisystem com email info hq digi co jp ...

Page 2: ...Page 2 of 6 Contents Overview 2 Features 2 Description of display 4 Key switch function 4 Function setting 5 Error messages 6 Operation mode 6 ...

Page 3: ... 20 times second 3 Internal resolution 1 300 000 4 Display resolution 1 1000 1 30 000 division 5 Non linearity 0 016 F S 6 Rated Output 0 20mV 7 Excitation Voltage _DC5V Analog up to 4 x 350Ω load cell 8 Power supply DC6V 4Ah re charge battery or AC230V power supply 9 Display 6 digits LED display and height 25mm LED 10 LED indicate G W N W tare zero ct oz lb kg 11 Power consumption 1W 12 Operating...

Page 4: ...t The weight is over the preset value OK Pass limit The weight I between HI LO preset value LQ Low limit The weight is under the preset value Switch bottom function 1 Switches Power On Off i Press I once the display show a self test and come to Zero on the screen Each LED will display 0 to 9 in sequence during the self test ii Press and hold I for 3 0 second the display will shut down automaticall...

Page 5: ... Press to choose On or OFF for buzzer sound or not sound 3 Press M the display show L XX X Press to select a new value for weight under the limit LO 4 Press M the display show 0 XX X Press to select a new value for weight between the HI LO limit OK 5 Press M the display show H XX X Press to select a new value for weight over the limit HI 6 Press M the display show 08 09 10 Mean is year month day 0...

Page 6: ... stable To check the flatform structure To open the platform structure and clean the structure The scale can not do calibration To check the calibration jumper put to ON position To open casing and palce the cal jumper ON position The scale is not accurate The Loadcell is not good in condition To replace the new Loadcell Operation Mode Open the casing switch the jumper JP1 to OFF position The func...