15
d.
--
The division will be displayed
P
x.y
The No. x parameter is set to y.
Lo.bAt
The voltage of batteries or input power is below 4.7V
9.
Trouble shooting:
SYMPTOM
PROBABLE CAUSE
REMEDY
Does not turn on.
1.
AC adapter is not securely connected
2.
Low battery
3.
Indicator is damaged
1.
Re-plug the AC adapter or rotate the plug to
securely connect to the scale
2.
Replace the batteries
3.
Replace
with
a new
indicator
&
perform
calibration
Ad
ˉ ˉ ˉ
1.
The cable from platform to indicator is
not correctly connected, or
disconnected, or shor t circuit
2.
Indicator is damaged
3.
Load cell cable is broken
4.
Load cell is damaged
1.
Check the cable or replace with a new cable
2.
Replace with a new indicator and perform
calibration.
3.
Return the scale for repair
Ad
_ _ _
0
ˉ ˉ ˉ
Indication is out of key zero range
Reduce weight on platter till indication is within key
zero range.
0
_ _ _
Weight reading below Power On Zero limit.
1.
Loosen the shipping protection screws
2.
Check whether an object is stuck between scale
base, if yes, remove the object.
3.
Perform zero calibration.
ˉ ˉ ˉ
1.
Weight reading exceeds Overload limit
2.
The weight value cannot be displayed in
the current unit of measure because it
exceeds 6 digits.
1.
Reduce load on the scale until a weight value is
displayed.
2.
Use a more appropriate unit of measure.
_ _ _
Weight reading below Under load limit
1.
loose the shipping protection screws
2.
Perform zero calibration
EEP.E1
The parameters are not same with backup
data
Re-set Setup parameters as technical manual
instructed.
EEP.E2
Setting parameter(s) is not in normal range
Re-set USER parameters per the Technical manual
CAL.Er
1.
Input data or loaded weight is too small,
too big
2.
Weight signal is unstable, un-linear
1.
Input correct data, load correct weight onto
platform.
2.
Return the scale for repair
Cannot zero the display
1.
Load on scale exceeds allowable limits.
(
20%FS
)
2.
Load on the scale is unstable
1.
Remove load from the scale.
2.
Wait for the load to stabilize. then press the
ZERO/ON/OFF key to zero the display
1. Max. CAPACITY not same
as marked on overlay
2. Any function invalid
3. Any units missed
CONFIG and USER parameters are not
correctly set
Re-set CONFIG and USER parameters per the
Technical manual
Weighing is not accurate
1.
An object is stuck between the load cell
and scale base
2.
Load cell received a heavy impact
1.
Remove the object
2.
Perform Linearity calibration
3.
Loose the shipping protection screws.
4.
Place the load on the center of the weighing
platform