10
Para-
meter
Default
Setting
PS-105-120
setting
P17.x
Default
Setting
PS-105-120
setting
1
Setting
PS-105-120
setting
Check weight stability range: 0=+0.5d; 1=±1d; 2=+1.5d; 3=+2d;
4=+3d; 5=+4d; 6=+5d; 7=+6d; 8=+7d; 9=+8d
18560120
Factory Setting Setting
PS-105-120
setting
1
P18.x
1
Overload limit range: 0=FS+0d; 1=FS+9d; 2=101%FS;
3=102%FS; 4=105%FS; 5=110%FS; 6=120%FS; 7=150%FS;
8=200%FS; 9=No limitation
1
P19.X
2
Backlight on-off mode selection (if it is installed):
0= Backlight is always off;
1= Backlight is always on;
2= Backlight is auto on and auto off. It is auto off after 10s when scale goes to
stable and has no key operation, and it is auto on when scale is unstable or
there’s some key operation.
2
4. Some Detailed Explanation
P3.x: RS232 mode setting
P3.0: No RS232 function. It will not transmit or receive any data although the scale is with RS232 hardware. RS232
function can be only activated when scale is in normal weighing mode.
P3.1: Continuously output of the current displayed reading and unit, and it does not receive any data.
The output format is as below:
<LF>< reading, minus, decimal point, weight unit><CR><EXT>
P3.2: Continuously output of the current gross weight, TARE weight and net weight reading data including unit, and
not accept any data. The format is as follows:
<LF><Gross: reading, minus, decimal point, unit><CR><EXT>
<LF> <TARE: reading, decimal point, unit><CR><EXT>
<LF> <Net: reading, minus, decimal point, unit><CR><EXT>
The number of position used: weight reading ---7; Minus ---1;
Decimal point ---1;
Weight unit ---2 or 4;
P3.3: When the scale is stable, it will output the current displayed weight reading automatically one time including
unit, and not accept data. The output format is same as P3.1
P3.4: When the scale is stable, it will output the current gross weight, TARE weight and net weight data including
unit automatically one time, and not accept data. The output format is same as P3.2
P3.5: Bio-RS232 data output.
P3.6: Continuously output of the current net weight in gram with five numbers. The output format is as below:
<LF>< x x x x x><CR><EXT>
If net weight is larger than range, “>>>>>”will be transmitted, and If net weight is below 0, “-----”will be
transmitted.
P3.7: When the scale is stable, it will output the current net weight in gram with five numbers. The output format is
same as P3.6.
P9.x: Calibration unit
As per the setting of P7, P8 and P9, following table is listed,