Why isn’t the chart keeping time or running slow?

•  The chart may be hung up or restricted, possibly caused by a rip on the outer edge of the chart or the CHART 

HUB, or the chart may be caught between the arm and pen arm platform. 

•  Incorrect chart installed for the selected chart speed. 

Why did the chart stop turning?

•  Chart hung up or restricted, (ripped chart) 
•  Unit may be “locked up”, this can be confirmed by pressing any of the buttons on the keypad, if the unit is locked 

up there will be no response to button presses and the Chart Recorder may appear to be working, but the readings 
won’t change, also the chart will not rotate. Remove power and battery, and then re-power. 

Why don’t the display and chart match?

•  DIP SWITCHES set for a specific range, but using chart for another range, or vice versa. 
•  Pen not inserted on pen arm all the way 
•  To adjust pen to match chart see “PEN HOME Adjust” section of this manual.

Why does the display read PROB?

•  Check the sensor pod and KTC probe. Make sure it is properly plugged into the unit.
•  Is the unit locked up? This can be confirmed by pressing any of the buttons on the keypad, if the unit is locked 

up there will be no response to button presses and the unit may appear to be working, but the readings won’t 
change, also the chart will not rotate. Remove power and battery, and then re-power.

Why does calibration seem to be off? 

•  What is the tolerance of the unit it’s being compared to? It’s ok if the unit is within the sum of the two tolerances. 
•  Has an outside calibration house attempted calibration? It may not have been adjusted properly. 
•  Walk through calibration adjustment, found in the manual

Why won’t the battery back-up work?

•  Are there good batteries in the chart recorder?
•  Keep in mind that battery backup varies significantly depending on temperature, pen movement and chart 

rotation speed. 

•  If there was a momentary power outage (Brown out), the unit may not have had enough time to recognize this 

and switch to battery mode. (The Chart Recorder may lock up or shut off in this situation) This situation may also 
occur if the unit is plugged into an outlet that shares the circuit with other machinery that has phase motors or 
compressors that cycle periodically. As these other machines cycle, they momentarily have a high current draw, 
therefore drawing from the unit.

Why won’t the chart recorder respond to changes?

•  Is the unit locked up? This can be confirmed by pressing any of the buttons on the keypad, if the unit is locked 

up there will be no response to button presses and the unit may appear to be working, but the readings won’t 
change, also the chart will not rotate. Remove power and batteries, and then re-power. 

Why won’t the unit power up?

•  Remove the batteries and power adapter for a minute or two, this will reset the unit. The unit should respond 

when the adapter is plugged in.

Dickson warrants that this line of instruments will be 
free from defects in material and workmanship under 
normal use and service for a period of twelve months 
after delivery.

This warranty does not cover routine calibration and 
battery replacement. For Specifications and Technical 
Support go to DicksonData.com.

Contact Customer Service at 630.543.3747 for a Return 
Authorization Number (RA) before returning any 
instrument. Please have the model and serial numbers 
and PO ready before calling.

Note: Dickson shall not be liable for consequential 
or incidental damages resulting from failure or 
malfunction of its products.

REPLACEABLE SENSOR™ CERTIFICATE 

OF VALIDATION

Dickson ensures that the Replaceable Sensors™ / Instruments listed below were 

developed, tested and validated together.

Replaceable Sensor™ Model: R400

Instrument Model: KT6P1, KT6P2, KT6P5 

Operation:

 The Instruments and Replaceable Sensor™ models listed above were 

specifically designed to work together. No other sensors will work with the Instrument 

models listed. No other Instrument will work with the Replaceable Sensors™ listed. 

The Replaceable Sensor™ must be connected to the Instrument for the Instrument to 

operate properly.

Calibration: 

Only the Replaceable Sensor™ is calibrated. The sensor, and all 

calibration defaults and adjustments are stored on the Replaceable Sensor™. Accuracy 

is strictly controlled by the sensor – no adjustments are made by the Instrument. 

Readings are sent from the Replaceable Sensor™ to the Instrument for storage 

purposes only. Certificates of Calibration / NISTs are only supplied for the Replaceable 

Sensor™ for this reason. 

Recalibration:

 When a Replaceable Sensor™ is due for recalibration the existing 

Replaceable Sensor™ can be replaced with a new Replaceable Sensor™ thus 

eliminating the need to return the entire instrument for recalibration. Replaceable 

Sensors™ can be returned for recalibration as well.

FAC

TROUBLESHOOTING

WARRANTY

FACTORY SERVICE AND RETURNS

6

7

Summary of Contents for KT6

Page 1: ...com Telephone 630 543 3747 Email support dicksondata com 930 S Westwood Avenue Addison IL 60101 KT6 6 TEMPERATURE CHART RECORDER USER MANUAL PRODUCT GUIDE BATTERY COMPARTMENT DIP SWITCHES ADJUST BUTTONS AC ADAPTER REPLACEABLE SENSOR PORT REFERENCE GUIDE ...

Page 2: ...r and that the Replaceable Sensor is plugged in firmly to the chart recorder 3 Insert 4AA batteries and plug in AC adapter Reference Guide on back Unit will power on 4 The instrument will move pen to the appropriate readings 5 Install the chart that matches the DIP SWITCH settings Reference settings label on the battery door to set the correct dip switch positions for the chart 6 Press the PEN HOM...

Page 3: ... display Note Make sure to press the ADJUST DOWN button first so the unit will not turn off 2 To raise the unit of measurement press the ADJUST DOWN button To lower the unit of measurement press the ADJUST UP button 3 When Calibration is complete press the ON OFF button The adjustment is stored in memory even afteryou turn the unit off or ifAC power fails Note Hold ADJUST UP orADJUST DOWN button f...

Page 4: ...from the unit Whywon t the chart recorder respond to changes Is the unit locked up This can be confirmed by pressing any of the buttons on the keypad if the unit is locked up there will be no response to button presses and the unit may appear to be working but the readings won t change also the chart will not rotate Remove power and batteries and then re power Whywon t the unit power up Remove the...

Reviews: