10. TROUBLESHOOTING
10.1. Regular Troubleshooting Procedures
From Maintenance submenu, user can initiate maintenance procedures such, cleaning,
priming, or draining chamber.
10.2. Weekly User Maintenance
Perform weekly maintenance before turning on the power switch. The right side has a side
door giving access to the fluidic system and the mechanical parts easily.
10.2.1. Cleaning needle washing head
Needle washing head cleans the outer surface of the aspirating needle with diluent.
Any salt build-up on the lower surface may cause malfunction during operation. Use a soft
cloth or wiper dampened with water to clean this area. You can see the washing head
indicated in the following figure:
Figure 12. Parts of measuring block
1. Exit Measure menu. Open the side door after the needle has stopped moving.
2. Gently rub the lower surface of the washing head with a damp cloth or wiper to remove
the salt build-up.
3. Close the side door.
Wash head
Measuring
chamber