ROUTINE UTILIZATION & MEASUREMENT
30
Adapter for 2 ml control blood
Figure 10. Vial used in control adapter
1. Remove the cap!! It is very important because the tip can not
pierce the cap!
2. Position the sample tube in the sample rotor.
3. Press START key.
The sample rotor will turn the vial into the inside of the instrument and needle draws
sample from the tube. The aspirating needle is retracted, while its outer surface is
automatically rinsed with diluent by needle wash head. This insures the low carry-
over between samples. After a few seconds, the rotor turns out. Now you can remove
the sample tube from the adapter.