SSE-

006 UV Water Sterilizer operator’s manual

 

 

SSE-006 UV 

water sterilizer Operator’s Manual

 version 04.07.19 

Page 

2

 / 

4

 

 

2.

 

Safety 

First of all, take care of your skin and your eyes from UV light !!! Never look directly at the light tube 

while device is in operation. Do not remove the UV Lamp from the Reactor Chamber when 

applying electrical power. Wear SAFETY GOGLES if you need to look necessary.  

 

Basic safety precaution should always be followed to reduce the risk of fire, electric shock. Make 
sure the leakage protection switch is installed before use. 

 
Do not touch the UV Lamp directly in case of burning. A pair of gloves would be better when taking 
off it.

 

3.

 

Precautions to do before using the device 

 
a.

 

Water Quality Guidelines 

The Ultraviolet Disinfection System is intended for the use with visually clear water, not colored, cloudy or 

turbid 

 

Ambient Water Temperature : 2-45°C 

 

Iron : < 0.3ppm (0.3mg/L) 

 

Hardness : < 7gpg (120mg/L) 

 

Turbidity : < INTU 

 

Manganese : 0.05ppm (0.05mg/L) 

 

UV Transmittance : > 75% 

b.

 

Recheck the installation before plus the sterilizer into power. 

c.

 

Do not proceed to install the equipment when UV Lamp or Sleeve Tube is broken, buy one again and 
continue. 

d.

 

Ultraviolet Disinfection System is designed for continuous operation and frequent switching will reduce 
Ultraviolet radiation and service life. But, if you do not use the equipment for a long time, you can turn 
off the power and water supply to save energy. 

e.

 

If this unit falls into the water, turn main power off and then retrieve it. Do not attempt to use this 
sterilizer if it has been submerged. 

f.

 

Do not operate this unit if it has damaged cord or plug, if it is malfunctioning or if it has been dropped or 
been damaged in any manner 

g.

 

The UV system should be installed after the filter on the return line 

h.

 

Always disconnect the water supply and completely drain the water purifier if it will be subjected to 
temperatures below freezing for extended periods of time. 

4.

 

Maintenance 

a.

 

Testing monthly or before each use 

b.

 

Lamp replacement is recommended every 8000 hours of operation. After 8000 hours, the lamp may still 
light, but the UV intensity has diminished. 

c.

 

Cleaning of the quartz sleeve once 3-6 months with alcohol or a mild detergent. 

5.

 

Spare parts 

6W UV Lamp reference :   959001 
SS006 UV lamp sterilizer:   959000 

 

 

Summary of Contents for SSE-006

Page 1: ...ons The assumption is made that before making an attempt to operate the device the operator is familiar with the operation of the water sterilizer and has read its Operator s Manual Importer DiaSysTechnologies 1682 rue de la Valsiere 34790 GRABELS FRANCE FieldService ContactyourlocalDiaSysDiagnosticrepresentativeforsupport DiaSys Technologies DiaSys Diagnostic Systems are registered trademarks of ...

Page 2: ...to power c Do not proceed to install the equipment when UV Lamp or Sleeve Tube is broken buy one again and continue d Ultraviolet Disinfection System is designed for continuous operation and frequent switching will reduce Ultraviolet radiation and service life But if you do not use the equipment for a long time you can turn off the power and water supply to save energy e If this unit falls into th...

Page 3: ...t size 1 4 female thread Voltage 220 240V 50 60Hz Lamp power 6w Max working pressure 8 bar Ambient water temperature 2 40 Visual audible lamp alarm YES Visual power on YES Lamp replacement reminder NO Lamp working life reminder NO UV LAMP Power 6w Diameter 15mm T5 type Installation available length 212mm Current 0 425A Voltage 20V uvc intensity 254nm 22uW cm2 Working life 8 000 hours 7 Content of ...

Page 4: ...ctor Chamber To protect the O ring do not over tighten Step 7 Connect 4 pin Electrical Socket with UV Lamp pins tightly Step 8 Carefully insert the UV Lamp into the Quartz Tube through End Nut Step 9 Install the Cover Cap and hand tighten onto End Nut Step 10 Select a readily accessible and well lit location to fix the system The system should always be located closest to the point of use and can ...

Reviews: