Diamond-MM-48-AT User Manual V1.01
Page 30
8.6 Convert the numerical data to a meaningful value
Once you have the A/D code, you need to convert it to a meaningful value. The first step is to
convert it back to the actual measured voltage. Afterwards you may need to convert the
voltage to some other engineering units (for example, the voltage may come from a
temperature sensor, and then you would need to convert the voltage to the corresponding
temperature according to the temperature sensor’s characteristics).
Since there are a large number of possible input devices, this secondary step is not included
here; only conversion to input voltage is described.
It is possible to combine both transformations into a single formula for computing efficiency.
To convert the A/D value to the corresponding input voltage, use the following formulas:
Conversion Formula for Bipolar Input Ranges
Input voltage = A/D value / 32768 * Full-scale voltage
Example: Input range is
±
10V and A/D value is 17761:
Input voltage = 17761 / 32768 * 10V = 5.420V
For a bipolar input range, 1 LSB = 1/32768 * Full-scale voltage. For the +/-10V range, 1 LSB
= 305
µ
V, and for the +/-5V range, 1 LSB = 153
µ
V.
Here is an illustration of the relationship between A/D code and input voltage for a bipolar
input range (V
FS
= Full scale input voltage, 10V or 5V):
A/D Code
Input voltage symbolic formula
Input voltage for
±
10V range
-32768
-V
FS
-10.0000V
-32767
-V
FS
+ 1 LSB
-9.9997V
. . .
. . .
. . .
-1
-1
LSB
-0.00032V
0
0
0.0000V
1
+1
LSB
0.00032V
. . .
. . .
. . .
32767
V
FS
- 1 LSB
9.9997V
Conversion Formula for Unipolar Input Range
Input voltage = (A/D value + 32768) / 65536 * Full-scale voltage
Example: Input range is 0-5V and A/D value is 17761:
Input voltage = (17761 + 32768) / 65536 * 5V = 3.855V
For a unipolar input range of 0-5V, 1 LSB = 1/65536 * 5V = 76
µ
V.
Here is an illustration of the relationship between A/D code and input voltage for a unipolar
input range (V
FS
= Full scale input voltage):
A/D Code
Input voltage symbolic formula
Input voltage for 0-5V range
-32768
0V
0.0000V
-32767
1 LSB (V
FS
/ 65536)
0.000076V
. . .
. . .
. . .
-1
V
FS
/ 2 - 1 LSB
2.4999V
0
V
FS
/ 2
2.5000V
1
V
FS
/ 2 + 1 LSB
2.5001V
. . .
. . .
. . .
32767
V
FS
- 1 LSB
4.9999V