9. SYMBOL INDICATING ARRHYTHMIA

Pro Afib pressure monitor is a fully automatic digital device that also analyses pulse 
rate  during  blood  pressure  measurement.    If  an  irregular  pulse  (arrhythmia)  is 

detected during measurement, the symbol (

 

) is displayed after completion of the 

measurement (when the measurement results are displayed). In most cases, this 
does not give cause for concern. However, if this symbol appears frequently (e.g. 
several times per week on measurements performed daily), we recommend you to 
inform  your  doctor.  This  device  does  not  replace  heart  diseases  diagnostics,  but 
serves as a device to detect arrhythmia at an early stage.

10. SYMBOL INDICATING  ATRIAL FIBRILLATION 

The  device  detects  atrial  fibrillation.  If  atrial  fibrillation  is  detected  during  the 

measurement, the symbol 

 

 is displayed after completion of the measurement 

(when  the  measurement  results  are  displayed).  If  this  symbol  is  displayed  after 
having performed a full measurement (two-fold measurement), we advise you to 
wait  one  hour  and  retake  your  measurement.  If  the  atrial  fibrillation  symbol  is 
displayed again, we recommend you to inform your doctor. 
If after repeated measurement, the symbol does not appear there is no cause for 
concern. In that case, it is advisable to take another measurement again the next day. 
The  device  may  not    detect  atrial  fibrillation  for  patients  with  pacemakers  and 
defibrillators.

11. ERROR MESSAGES

If  an  error  occurs  during  the  measurement,  the  reading  will  be  interrupted

  and  an  error  code 

displayed.

PAGE 10

Error code

 

Possible cause ERR 

ERR 1 

No pulse detected.

ERR 2  

Measurement results affected by interference. 

Cause: there was an arm movement during measurement.

ERR 3  

Inflation of the cuff has taken too long. The cuff has not been wrapped properly.

ERR 5  

Measurement has indicated unacceptable difference between the systolic and 

diastolic pressure values. Perform another measurement carefully following the 
instructions. If unusual results persist, contact your doctor.

Further information. Blood pressure varies even in healthy people, that is why it is important to 
always  take  measurements  under  the  same  conditions  (peaceful  environment).  If,  despite 
following these principles, the fluctuations will be higher than 15 mmHg and irregular pulse rate 
occurs  repeatedly,  consult  your  doctor.  In  the  event  of  problems,  you  should  consult  with 

Diagnosis S.A. 
YOU SHOULD NEVER ATTEMPT TO REPAIR THE DEVICE YOURSELF! ALL UNAUTHORIZED 
ATTEMPTS AT OPENING THE DEVICE WILL VOID THE WARRANTY!

 

If, during the use of the device, a problem occurs, please check the following items and undertake 
the listed remedial measures.

Summary of Contents for PRO Afib

Page 1: ...STRUCTION MANUAL AUTOMATIC UPPER ARM BLOOD PRESSURE AND PULSE MONITOR WITH ATRIAL FIBRILLATION DETECTION Diagnosis S A ul Gen W Andersa 38A 15 113 Bia ystok Poland tel fax 85 732 46 22 732 40 99 www d...

Page 2: ...GN 4 5 STARTING THE DEVICE 5 5 1 Installing battery 5 5 2 Battery life 5 5 3 Power supply 5 5 4 User selection and setting date and time 6 6 TAKING MEASUREMENTS 7 6 1 Before measurement 7 6 2 Most fre...

Page 3: ...ubing because this may result in a serious injury of the patient or disturbances in blood pressure measurement l Too frequent measurements may cause trauma to the patient due to impaired blood flow l...

Page 4: ...lic pressure falls below 60 mmHg you should also consult your doctor Even in the case of pressure values in the normal range it is recommended to perform regular blood pressure self measurements This...

Page 5: ...wever for people from the age of 50 years with high blood pressure hypertension diabetes coronary heart failure or have had a previous stroke atrial fibrillation screening is also recommended Early di...

Page 6: ...EMENTS LCD display MEMORY button power supply socket TIME button air tube connector diastolic pressure systolic pressure date time exhausted battery symbol blood pressure classification indicator acco...

Page 7: ...which they are used and may be shorter at lower temperatures If the low battery symbol is displayed they should be replaced with new ones 5 3 Power adapter 1 Connect the plug of the power cord into t...

Page 8: ...display the date and time Now the entered settings are confirmed and the clock starts running After pressing the TIME and MEMORY buttons data is entered e g switching from hours to minutes or changin...

Page 9: ...easurement Relax rest the elbow on the table with palm facing up the cuff should be at heart level Accuracy of readings may be reduced if the cuff is not wrapped properly The arm should be at the same...

Page 10: ...a 15 second interval between blood pressure readings is accurate in accordance with Blood Pressure Monitoring 2001 6 145 147 A timer counts down showing how much time remains for starting a new measur...

Page 11: ...ton for at least 5 seconds Release the button when the screen displays CL To permanently delete the entire memory press the MEMORY button while the CL is flashing PAGE 9 EXAMPLE 5 Systolic pressure 12...

Page 12: ...concern In that case it is advisable to take another measurement again the next day The device may not detect atrial fibrillation for patients with pacemakers and defibrillators 11 ERROR MESSAGES If a...

Page 13: ...fter it has been properly wrapped l The device is not intedud for usa in the electromagnetic enviranment generated by all phones or radio l Do not use the device if it is damaged l If the device is no...

Page 14: ...and the environment which may be threatened by inappropriate handling of waste If you are in doubt where to return the used appliance contact Diagnosis Free infoline 800 70 30 11 Measurement Display m...

Page 15: ...kV for 1 kV differential mode 2 kV common mode 5 UT 95 clip in UT for 0 5 cycle 40 UT 60 clip in UT for 5 cycle 70 UT 30 clip in UT for 25 cycle 5 UT 95 dip de UT dla 5 s 3 A m Compatibility 6 kV con...

Page 16: ...sion and TV transmission cannot be predicted theoretically with accuracy To assess the electromagnetic environment tests of local conditions should be considered If the measured field strength in the...

Page 17: ...inconsistent with the user manual ingress of liquids or foreign bodies overvoltage of mains repairs by unauthorized persons and random events 4 Faulty equipment should be delivered by the buyer to the...

Page 18: ...w Polska tel 85 874 60 45 serwis diagnosis pl Diagnosis S A ul Gen W Andersa 38A 15 113 Bia ystok Poland Free infoline 800 70 30 11 tel fax 85 732 46 22 732 40 99 www diagnosis pl Rev 2016 09 30 Angie...

Reviews: