DOUBLE HEAD PERISTALTIC PUMP
USER MANUAL
Put sturdy tubing in plastic cassette: turn each cassette counter of clockwise until they stops then pull them out. Add
grease to the cassette shafts and the surface of new tubing. Put a piece of sturdy tubing in (between rotor and cassette
as shown on picture below) Place back the cassette to the base. Do the same for the other pump head.
Cut with scissors the silicon tube standing for the warm and cold flow where you want to integrate the pump.
Then connect each silicon tube ending to the appropriate
sturdy tube ending.
Watch direction of flow: each pump has a drawing side and
a discharging side. Heads spin clockwise which means left
nozzles are the discharging side and right ones are the drawing
side. Have a look on the global scheme page 7 to make sure.
Place plastic pipe holders to fix pipes along.
Concerning the upper lid of the Bioruptor’s bath, set two plastic
nozzles over the upper part of the sonication bath. Press firmly
to ensure fixation.
Option 1:
- Connect the warm flow pipe and the cold flow pipe from the sonication water reservoir to the chiller water reservoir.
Grease up
Rotor
Knot is installed inside
of the guide
Plastic tube holders
Upper part of the sonication bath