background image

PU690 - Proving Unit

690V AC Proving for Voltage 

Indicators

Instruction Manual

Summary of Contents for PU690

Page 1: ...PU690 Proving Unit 690V AC Proving for Voltage Indicators Instruction Manual ...

Page 2: ...asurably better Contents I Overview 3 II Safety Notice 3 III Instrument Symbols 4 5 IV Technical Specification 6 V Display and Controls 7 1 Operating Instructions 7 8 2 Battery Replacement 8 3 Accessories 9 4 Warranty Information 10 1 ...

Page 3: ...g Unit to which this declaration relates is in conformity with the relevant clauses of the following standards EN 61010 1 EN 61010 2 030 EN 61326 1 EN 61326 2 2 LVD EMC RoHS The safety and performance of this instrument is assured when operated within the specifications in this instruction manual The product identified above conforms to the requirements of council directive 2014 30 EU and 2014 35 ...

Page 4: ...the warnings and operating instructions before use If the instrument is used in a manner not specified by the manufacturer the protection provided by the instrument may be impaired 2 If the proving unit appears damaged or is not working properly stop using it immediately The operator s safety is no longer ensured the instrument is to be put out of service and protected against use Safety may be co...

Page 5: ...ge Directive 2014 35 EU Standards EN 61010 1 EN 61010 2 030 and EN61326 1 are fulfilled III Instrument Manual Symbols Symbols displayed on the instrument and in the instruction manual II Safety Notices continued 4 Do not operate the proving unit for periods longer than 10 seconds 3 Clean the tester casing with a damp cloth and mild detergent Do not use abrasives or solvents 4 When low battery indi...

Page 6: ... where the product was purchased They can take this product for environmental safe recycling The instruction manual contains information and references necessary for safe operation and maintenance of the instrument Prior to using the instrument the user is kindly requested to thoroughly read the instruction manual and comply with it in all sections Failure to read the instruction manual or to foll...

Page 7: ...750V Operating temperature 10 C to 40 C at max Storage temperature 10 C 50 C Relative humidity 70 non condensing Altitude Up to 2000m Pollution Degree 2 Dimensions H W D 140mm x 80mm x 50mm Weight net 400g Power supply 6 x 1 5V LR6 AA supplied EAN 5060082543562 Measurement category II is for measurements performed on circuits directly connected to the low voltage installation Examples of this woul...

Page 8: ...on the voltage tester under test Ensure the output voltage LED is on during the test Once visual indication has been confirmed and verified remove the right test probe first before removing the left test probe 7 Please note the output voltage may not be 690V and will depend upon the load applied by the voltage indicator V Display and Controls unit Proving PU690 690V AC Output Voltage On LED Voltag...

Page 9: ... battery compartment cover Remove the battery compartment cover Remove the discharged batteries from the compartment Fit a new set of LR6 AA 1 5v batteries Relocate the battery cover over the battery compartment and fasten in position with the battery cover captive screw 8 All Di LOG voltage indicators draw 3 5mA and will indicate up to 690V but other manufacturers instruments may vary ...

Page 10: ...PU690 Operating Manual Di LOG measurably better 9 3 Accessories Optional Accessories for PU690 CCDLPK Combined carry pouch for PU690 Proving Unit and CombiVolt Voltage Tester User Notes ...

Page 11: ...proof of purchase Faults in manufacture and material defects will be rectified by us free of charge provided the instrument has not been tampered with and returned to us unopened Damage due to dropping abuse or misuse are not covered by the warranty Product Support support dilog co uk Di LOG Ltd Unit 28 Wheel Forge Way Trafford Park Manchester M17 1EH UK Tel 44 0 161 877 0322 Email sales dilog co ...

Page 12: ......

Reviews: