Final inspection
Inspect Castor for signs of wear and damage. Ensure that Castor stands stable on the floor, so that it does not tilt,
and check that the uprights are adjusted to the same height.
Check to ensure that all locking handles and bolts are securely tightened.
Test lift function by lifting a person (not a patient) using an approved sling.
Keep the instruction manual where it is accessible to users of the product.
Safe working load
Different products on the same lift system (lift unit, slingbar, sling, scales and other lifting accessories) may have different
allowable safe working loads. The lowest allowable safe working load always determines the maximum load of the
assembled system. Always check the safe working loads for the lift and accessories before use. Contact your dealer if
you have any questions.
Important information
• Castor must be assembled according to the assembly instructions provided with the stand.
• Castor may only be used indoors and on a level floor.
• Lifting accessories must be properly trial fitted and tested in relation to the patient’s needs and functional ability.
• Do not leave the patient unattended during a transfer situation.
• Under no circumstances may max. load be exceeded. See section on safe working load.
• To ensure optimal function, Castor must be inspected regularly. See section on maintenance.
• Warranty applies only if repairs or alterations are done by an authorized technician.