PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 84
Table 13.
Reset – Cal
RESET
RESULT
SEE SECTION
All Q-RPT user calibration
coefficients
PA to zero, PM to 1.
5.3.1.1, 5.3.7
Q-RPT absolute mode
ON.
5.3.5
Utility Sensor calibration
coefficients
PA to zero, PM to 1.
5.5
On-board barometer
calibration coefficients
PA to zero, PM to 1.
5.4
Calibration date
Set all dates to 19800101.
5.3.7, 5.5, 5.4
AutoZ values
All P
offset
values to zero for absolute mode and
101325 Pa for gauge modes.
3.5.1
AutoZ function
ON, all Q-RPTs, all measurement modes.
3.5.1
3.5.4.5
<5ALL>
PURPOSE
The reset - all function clears and deletes large amounts of user defined
information including critical calibration data.
Combines all resets in one global reset command that clears the entire user
section of non-volatile memory returning PPCH to the “as delivered” condition
except the ID function (see Section
3.5.5.4) and the security level password (see
Section
3.5.5.5). The Reset – All resets are itemized in Table 14.
Table 14.
Reset – All
RESET
RESULT
SEE SECTION
Reset – Sets
All the resets of Reset – Sets
3.5.4.1
Reset – Units
All the resets of Reset – Units
3.5.4.2
Reset – Atest
All the resets of Reset – Atest
3.5.4.3
Reset – Cal
All the resets of Reset – Cal
3.5.4.4
Remote
Interfaces
COM1, COM2 and IEEE-R88 interfaces to
default settings.
3.5.2
Remote Communications
Remote command format to Classic.
3.5.2.3
Level (Security)
Low.
3.5.5.5
Valve
Config
Return to factory valve opening points to
factory values.
3.5.7.1
3.5.5
<5PREF>
PURPOSE
To access a menu of PPCH operational preferences and functions.
OPERATION
To access the PREF menu press
[SPECIAL]
,
<5pref>
. The display is:
1ScrSvr 2sound 3time
4ID 5level