MOLSTIC-S™ OPERATION AND MAINTENANCE MANUAL 

 

© 2004-2007 DH Instruments, a Fluke Company

 Page 

24

 

2.4.3

 

INLET CONNECTION - MOLSTIC-S USE WITH DUT 
DOWNSTREAM 

  Connect the Gas Supply: 

Connect the gas supply to the inlet fitting of the molstic-S 

using suitable means and compatible fittings.  Following instructions in Section 2.4.1.

 

2.4.4

 

OUTLET CONNECTION - MOLBLOC-S OUTLET 
CONNECTION 

The outlet fitting of the molbloc-S utilizes the ISO-KF Style vacuum flange. Use the following 
procedure for connecting and disconnecting molbloc-S elements with ISO-KF Style connections. 

 

 

 

1. 

Flange, ISO-KF style, 16 mm 

Figure 19.  

molbloc-S with ISO-KF Style Flange

 

 

2.5

 

CONNECTING, DISCONNECTING AND MOUNTING 
MOLBLOC-S

 WITH ISO-KF STYLE VACUUM FLANGE 

(MOLBLOC-S OUTLET FITTING) 

2.5.1

 

ISO-KF STYLE VACUUM FLANGE FITTINGS

 

The outlet connection system on the molbloc-S body is the ISO-KF style vacuum flange.  It 
utilizes an internal centering ring, an external overpressure ring and clamp.  The seal is 
effected by the uniform application of pressure by the clamp on the 15° surface of the mating 
stainless steel flanges.  These mating flange surfaces compress a Viton O-ring that is held in 
place by the centering ring.  The overpressure ring keeps the O-ring in place, and maintains a 
leak free connection when the system is subjected to internal pressures above vacuum level.  
This connection is reusable, rotatable, and can operate leak free in vacuum applications up to 
10

-8

 Torr, and in positive pressures of 700 kPa (100 psig). 

 

 

When performing a leak test of a plumbing system that contains the ISO-KF Style 
Vacuum Flanges, do not exceed the maximum operating pressure of the molbox:   

600 kPaa (87 psia) for molbox1-A700K, 250 kPaa (36 psia) for molbox1-A350K models.

 

 

Summary of Contents for MOLSTIC-S

Page 1: ...2004 2007 DH Instruments a Fluke Company molstic S Engineered Mounting Systems for molbloc S Mass Flow Elements Operation and Maintenance Manual...

Page 2: ...warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual or produ...

Page 3: ...STIC S DUAL 1 4 IN WITH LOW OR MID FLOW METERING VALVE AND REGULATION 13 2 INSTALLATION 15 2 1 UNPACKING AND INSPECTION 15 2 1 1 REMOVING FROM PACKAGING 15 2 1 2 INSPECTING CONTENTS 15 2 1 2 1 ACCESSO...

Page 4: ...EGULATOR MAXIMUM PRESSURE ADJUSTMENT 31 3 2 3 LOCKING THE BACK PRESSURE REGULATOR ADJUSTMENT OPTIONAL 31 3 2 4 BACK PRESSURE REGULATOR OPERATION 31 3 3 LEAK CHECK RECOMMENDATIONS 32 3 4 MOLSTIC S FLOW...

Page 5: ...ure 7 molstic S Dual 1 4 in with Regulation 8 Figure 8 molstic S Dual 1 2 in with Regulation 9 Figure 9 molstic S Single 1 4 in with Low or Mid Flow Metering Valve 10 Figure 10 molstic S Dual 1 4 in w...

Page 6: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Page 7: ...rovide pressure control capability to the user Calibration test devices that are capable of controlling flow must be installed upstream of the molbloc S if they are to be tested in control mode Use th...

Page 8: ...ORDERING INFO DESCRIPTION P N FAM0009 molstic S molstic for molbloc S Option 01 1 Single Channel 1 4 in Plumbing Figure 1 molstic S Single 1 4 in Standard Configuration 1 Inlet adaptor 1 4 in tube x...

Page 9: ...09 molstic S molstic for molbloc S Option 01 2 Single Channel 1 2 in Plumbing Figure 2 molstic S Single 1 2 in Standard Configuration 1 Inlet adaptor 1 2 in tube x 1 2 in male VCR 2 Flow shut off mete...

Page 10: ...molstic S molstic for molbloc S Option 01 3 Dual Channel 1 4 in Plumbing Figure 3 molstic S Dual 1 4 in Standard Configuration 1 Inlet adaptor 1 4 in tube x 1 4 in male VCR 2 Particulate filter 2 mic...

Page 11: ...olstic S molstic for molbloc S Option 01 4 Dual Channel 1 2 in Plumbing Figure 4 molstic S Dual 1 2 in Standard Configuration 1 Inlet adaptor 1 2 in tube x 1 2 in male VCR 2 Flow shut off flow meterin...

Page 12: ...3 2 With 1 4 in Pressure Reducing Regulator 03 4 With 1 4 in Back Pressure Regulator Figure 5 molstic S Single 1 4 in with Regulation 1 Inlet quick connect includes stem with 1 4 in SWG not shown 2 Pr...

Page 13: ...03 3 With 1 2 in Pressure Reducing Regulator 03 5 With 1 2 in Back Pressure Regulator Figure 6 molstic S Single 1 2 in with Regulation 1 Inlet quick connect includes stem with 1 2 in SWG not shown 2 P...

Page 14: ...With 1 4 in Pressure Reducing Regulator 03 4 With 1 4 in Back Pressure Regulator Figure 7 molstic S Dual 1 4 in with Regulation 1 Inlet quick connect includes stem with 1 4 in SWG not shown 2 Pressur...

Page 15: ...3 With 1 2 in Pressure Reducing Regulator 03 5 With 1 2 in Back Pressure Regulator Figure 8 molstic S Dual 1 2 in with Regulation 1 Inlet quick connect includes stem with 1 2 in SWG not shown 2 Pressu...

Page 16: ...4 in Plumbing 04 2 With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 9 molstic S Single 1 4 in with Low or Mid Flow Metering Valve 1 Inlet adaptor 1 4 in tube...

Page 17: ...With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 10 molstic S Dual 1 4 in with Low or Mid Flow Metering Valve 1 Inlet adaptor 1 4 in tube x 1 4 in male VCR 2...

Page 18: ...Pressure Regulator 04 2 With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 11 molstic S Single 1 4 in with Low or Mid Flow Metering Valve and Regulation 1 Inlet...

Page 19: ...r 04 2 With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 12 molstic S Dual 1 4 in with Low or Mid Flow Metering Valve and Regulation 1 Inlet quick connect incl...

Page 20: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...

Page 21: ...2 1 ACCESSORIES INCLUDED WITH MOLSTIC S SHIPMENTS Table 1 Accessory Kit 1 4 in molstic S versions P N 401938 DESCRIPTION PART 1 molstic S Operation and Maintenance Manual 550130 1 16 mm ISO KF Blanki...

Page 22: ...ility such as MFCs and are to be tested in flow control mode must be tested in the upstream location Install this type of DUT upstream of the molbloc S only 2 3 2 INSTALLATION AND SWAPPING OF MOLBLOC...

Page 23: ...22 for use with the new flange Be careful not to damage this seal in any way as a leak can result which will cause errors during flow measurement Position the replacement VCR inlet flange on the end o...

Page 24: ...sconnecting molbloc S elements with VCR connections Connecting Disconnecting and Mounting molbloc S with VCR Face Seal Fittings molbloc S Inlet Fitting In making the molbloc S inlet flow connection al...

Page 25: ...cked off Never use wrenches to hold the molbloc S body 2 3 2 2 2 molbloc S Connections to molbox For the pressure connections use the molbox to molbloc S pressure tubes P N 401125 supplied with the mo...

Page 26: ...n the DUT and the molbloc The gas supply to the molstic S has the following requirements The gas supply must be clean and free of particulate contamination in order to avoid damage to the molbloc For...

Page 27: ...ade kits that are available in 1 4 in or 1 2 in P N 401880 and 401881 respectively Figure 16 molstic S with Supplied Regulator DHI also offers the supply only platform that includes the regulator with...

Page 28: ...tem leak testing The pressure reducing regulators supplied by DHI are selected for the following characteristics High quality High sensitivity and stability Outlet control pressure range 0 to 100 psig...

Page 29: ...om particulate contamination Consult factory for availability of 1 2 in inline filter kit Connect DUT to molstic S inlet Since there are so many types and sizes DUTs that can be calibrated by the molb...

Page 30: ...TYLE VACUUM FLANGE FITTINGS The outlet connection system on the molbloc S body is the ISO KF style vacuum flange It utilizes an internal centering ring an external overpressure ring and clamp The seal...

Page 31: ...e separated Remove the clamp centering ring and overpressure ring Never use wrenches to hold the molbloc S body or tighten the wing nut 2 5 2 1 MOLSTIC S USE WITH DUT UPSTREAM During operation the mol...

Page 32: ...he Downstream Tee Assembly end fitting type is the 25 mm ISO KF style vacuum flange These fittings can be converted to either Swagelok tube fitting or VCR Face Seal fittings using one of the following...

Page 33: ...as not designed to accommodate the DUT directly on its platform For the sake of flexibility the molstic S is supplied with 50 cm of PFA tubing to help make the inlet and outlet connections Use of the...

Page 34: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 28 N NO OT TE ES S...

Page 35: ...er to obtain the most accurate and stable set point 3 1 1 SYSTEM PRESSURIZAION Begin with system pressure vented Make sure that the system pressure is vented Set inlet pressure to minimum Turn the reg...

Page 36: ...sure Vent the system pressure Use the procedure in Section 3 1 to re pressurize and verify the regulator pressure setting Readjust if necessary Replace regulator knob cover Place the regulator adjustm...

Page 37: ...te the back pressure regulator adjustment knob clockwise in order to raise the pressure level upstream of the regulator Rotate the back pressure regulator adjustment knob counterclockwise in order to...

Page 38: ...S and will provide the control of inlet pressure to the molbloc S If the DUT does not have its own flow control capability the inlet pressure to the molbloc S must be controlled by either an upstream...

Page 39: ...the test system Adjust system flow rate Rotate the flow control valve adjustment knob to set the desired flow rate on the molbloc S Fine adjustments in flow rate can be achieved by combining adjustme...

Page 40: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 34 N NO OT TE ES S...

Page 41: ...com 4 2 UPGRADE OF MOLSTIC S WITH 1 4 IN METERING VALVE The 1 4 in molstic S is delivered with integrated shut off metering valves These valves provide a measure of flow control when calibrating devi...

Page 42: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 36 N NO OT TE ES S...

Page 43: ...stic S filter is a sealed 2 micron pleated mesh filter Swagelok SS 4FW VCR 2 It can be cleaned by backflushing with clean gas but must be removed from the molstic S When removing cleaning and re insta...

Page 44: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 38 N NO OT TE ES S...

Reviews: