GPC1-10000 / GPC1-16000
™
OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 2
1.2
SPECIFICATIONS
Electrical Power Requirements
None
Pneumatic Power Requirements
Clean, dry, low flow, compressed gas. Maximum pressure
needed depends on maximum pressure to be adjusted. Max
flow 200 sccm:
Working P
Drive Air P
- For 5 MPa (750 psi):
300 kPa (45 psi)
- For 50 MPa (7 500 psi):
600 kPa (87 psi)
- For 70 MPa (10 000 psi)
700 kPa (100 psi)
- For 110 MPa (16 000 psi):
850 kPa (120 psi)
Operating Temperature Range
10 to 35 °C
Weight
24 kg (53 lb)
Dimensions
21.5 cm H x 30 cm W x 53.5 cm D (8.5 in. x 11.8 in. x 21.1 in.)
Pressure Range
GPC1-16000: 0 to 110 MPa (16 000 psi)
GPC1-10000: 0 to 70 MPa (10 000 psi)
Maximum adjustable pressure depends on pneumatic power
supply (see above).
Operating Medium
Any non-corrosive gas
Pneumatic Power Connection(s)
1/4 in. NPT F
Test Connections
(3) DH500 F test connections ((1) left rear side and (2) on top)
DH500™ is a gland and collar type fitting for 1/4 in.
(6.35 mm) coned and left hand threaded tube. DH500 is
equivalent to AE F250C, HIP HF4, etc.
Pneumatically Actuated
Variable Volume (PDVV)
Displacement
2 cc (0.12 in
3
)