GB-K Gas Booster Package Operation and Maintenance Manual
© 1998 - 2000
DH Instruments, Inc
.
Page 2
PNEUMATIC CONNECTION PANEL:
1.
High Pressure Outlet Fitting:
a stainless steel DH 200 female coned end fitting used to
connect the high pressure output to a point-of-use (usually the PPCK+ SUPPLY port).
DH200 F is a gland and collar type fitting for coned and left hand threaded 1/4 in. (6.35 mm) OD
tube, equivalent to AE SF250C, HIP LF4, etc. DO NOT USE INCORRECT FITTINGS; DAMAGE TO
THE FITTING AND DANGER TO THE OPERATOR COULD RESULT.
2.
Low Pressure Outlet Fitting:
a stainless steel 1/8 in. NPT female fitting used to connect the low
pressure output to a point-of-use (usually the PPCK+ DRIVE IN port).
3.
Supply Inlet Fitting:
a stainless steel 1/8 in. NPT female fitting used to connect an external
source of high pressure gas. This is the gas pressure that will be boosted to obtain very high
pressure at the High Pressure Outlet and regulated to provide low pressure at the Low
Pressure Outlet.
Figure 2.
Control Panel
8
7
9
10
11
12
4
5
6
13