GB-H-152 GAS BOOSTER PACKAGE OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 4
1.4
SPECIFICATIONS
Dimensions
50 cm (20 in.) H x 57 (22.5 in.) W x 30 (12 in.) D
Weight (shipping)
44.4 kg (98 lb)
Drive air pressure supply
Range
Flow
0.15 to 1 MPa (20 to 150 psi)
425 to 2 125 slm (15 to 75 scfm)
Test gas pressure supply
Range
Flow
4.8 to 40 MPa (700 to 6 000 psi)
140 to 560 slm (5 to 20 scfm)
Output pressure
GB-H-152-70M
GB-H-152-100M
0.70 to 100 MPa (400 to 15 000 psi)
0.35 to 120 MPa (200 to 17 400 psi)
Operating media
air, helium, nitrogen
Pressure connections
Drive air supply
Shop drive out
Test gas pressure supply
High pressure gas out
1/4 in. NPT F
1/4 in. Swage
1/8 in. NPT F
DH500F
(
DH500 is a gland and collar type fitting for ¼ in. (6.35 mm)
coned and left hand threaded tube. DH500 is equivalent to AE
F250C, HIP HF4, etc.)
Nominal gas booster piston ratio
High pressure accumulator volume
152:1
150 cc (9.2 in.
3
)
Due to a policy of continual product improvement, all specifications are subject to change without notice.