background image

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL 

©1998-2008 DH Instruments, a Fluke Company 

Page 12

 

2.4

 

INITIAL START UP 

After installing the gas booster (see Section 2.3.2), perform the initial start up as follows: 

n

  Supply drive air:  Check that the Drive Air Control Kit 

SHUTOFF valve

 is in the 

OFF

 position and  

that the 

DRIVE ADJUST regulator

 is set to zero (fully counterclockwise).  Apply drive pressure to the 

SHUTOFF valve

 (see Section 2.2.2.1 for information on drive air supply requirements). 

o

  Supply high pressure instrument gas:  Check that all connections to the high pressure 

IN 

and 

OUT 

ports 

are secure.  Apply high pressure gas to the high pressure 

IN port 

(see Section 2.2.2.2 for 

information on high pressure instrument gas). 

 

  Ensure the high pressure supply does not exceed 20 MPa (3 000 psi).  Pressures above this range may 

result in a failure that could damage the instrument and/or cause personal injury. 

p

 Adjust the 

DRIVE ADJUST regulator

:  Rotate the 

DRIVE ADJUST regulator

 knob (clockwise) 

while observing the set pressure on the 

DRIVE gauge

.  Set the drive pressure to the maximum 

desired high pressure output divided by the booster ratio (75 for GB-75, 152 for GB-152).   
For example, if the desired high pressure output of a GB-152 is 70 MPa (10 000 psi), set the drive air 
pressure to 70 MPa/152 = 460 kPa (67 psi).  Remember that when the 

SHUTOFF valve 

is in the 

ON 

position, the booster will operate until the high pressure output is equal to the drive air setting times 
the booster ratio. 

 

  The DRIVE ADJUST regulator is NOT self venting so it cannot be used to set pressure lower than the 

current pressure when the SHUTOFF valve is OFF.  If you overshoot the desired set point, set the SHUTOFF 
valve to ON until the DRIVE pressure is lower than your desired setting.  Then set the SHUTOFF valve to OFF 
and use the regulator to adjust the pressure in the ascending direction. 

q

  Generate high pressure:  Set the 

SHUTOFF valve 

to the 

ON 

position.  The booster will begin to cycle 

and will continue to cycle until it stalls.  When it stalls, the high pressure output is roughly equal to the 
drive pressure setting times the booster's ratio (75 or 152).  

Summary of Contents for GB-152

Page 1: ...1998 2008 DH Instruments a Fluke Company GB 75 and GB 152 Gas Booster With Control Kit Operation and Maintenance Manual...

Page 2: ...ruments a Fluke Company All rights reserved Information in this document is subject to change without notice No part of this document may be reproduced or transmitted in any form or by any means elect...

Page 3: ...SITE 8 2 2 2 GAS SUPPLIES 8 2 2 2 1 DRIVE AIR SUPPLY 8 2 2 2 2 HIGH PRESSURE INSTRUMENT GAS 9 2 3 INSTALLATION AND SETUP 9 2 3 1 GENERAL CONSIDERATIONS 9 2 3 2 INSTALLATION 10 2 3 2 1 BOOSTER INSTALLA...

Page 4: ...NERAL INFORMATION 17 5 2 BOOSTER WILL NOT RUN 17 5 3 BOOSTER RUNS TOO SLOWLY 18 5 4 PRESSURE GENERATES TOO SLOWLY OR NOT AT ALL 18 5 5 BOOSTER RUNS CYCLES CONTINUOUSLY 19 5 6 CANNOT ACHIEVE DESIRED PR...

Page 5: ...Instruments a Fluke Company T TA AB BL LE ES S Table 1 GB 75 or GB 152 Parts List 7 Table 2 DHI Authorized Service Providers 21 F FI IG GU UR RE ES S Figure 1 GB 152 Gas Booster 2 Figure 2 Drive Air C...

Page 6: ...Fluke Company Page IV A AB BO OU UT T T TH HI IS S M MA AN NU UA AL L Manual Conventions CAUTION is used in throughout the manual to identify user warnings and cautions NOTE is used throughout the man...

Page 7: ...0 psi minimum to higher pressure The booster is powered by drive air The high pressure output is equal to the drive air input times the boosting ratio of 75 or 152 The high pressure output is connecte...

Page 8: ...OF THE COMPONENTS 1 2 1 GAS BOOSTER 1 High Pressure Instrument Gas Filter 2 High Pressure Instrument Gas Supply Port 3 Drive Air Supply Port from Drive Air Control Kit 4 Breather Port 5 Low Pressure...

Page 9: ...Company 1 2 2 DRIVE AIR CONTROL KIT 1 Drive Air Supply Connection not visible 2 Drive Pressure Adjust Regulator 3 Drive Pressure Indication Gauge 4 Drive Air Shut Off Valve 5 Drive Air Out Connection...

Page 10: ...Instruments a Fluke Company Page 4 Figure 3 System Schematic 1 Drive Air Regulator 5 High Pressure Outlet Port 2 Drive Air Gauge 6 Booster Pump 3 Drive Air Shutoff Valve 7 1 4 in PFA Tubing 4 High Pr...

Page 11: ...pressure output is high pressure supply times 25 Supply Flow Rates Shop Drive Air High Pressure Supply Gas 425 to 2 125 slm 15 to 75 scfm 140 to 560 slm 5 to 20 scfm Pressure Connections Drive Air Su...

Page 12: ...GB 75 AND GB 152 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 6 N NO OT TE ES S...

Page 13: ...spect all parts for damage If damage is noted report it to your Receiving Department for appropriate action Inspect for any missing components or accessories using Table 1 Should any items be missing...

Page 14: ...ion 2 3 1 for additional recommendations 2 2 2 GAS SUPPLIES Two sources of compressed gas are required to operate the gas booster They are drive air see Section 2 2 2 1 and high pressure instrument ga...

Page 15: ...valve turns the booster ON and OFF The Drive Air Control Kit is presented on a bracket meant to be mounted at a location convenient to the operator for example on the front of a work bench When select...

Page 16: ...steps below n Identify an appropriate location for the Drive Air Control Kit bracket If the booster will regularly be turned ON OFF and or frequent adjustment of booster output pressure is desired th...

Page 17: ...as the DHI GPC1 Gas Pressure Controller The booster high pressure OUT port is connected to the SUPPLY port of the GPC1 see the GPC1 Operation and Maintenance Manual To connect the booster high pressu...

Page 18: ...UST regulator knob clockwise while observing the set pressure on the DRIVE gauge Set the drive pressure to the maximum desired high pressure output divided by the booster ratio 75 for GB 75 152 for GB...

Page 19: ...air cycling valve pilot valve and vent section This section provides the reciprocating action and compression force needed to operate the booster and generate the high pressure gas Drive air is chann...

Page 20: ...imes the booster ratio To correctly set the pressure gas flow must not occur If flow is present in the circuit when the regulator is adjusted the pressure will increase when flow is reduced If the set...

Page 21: ...age 15 1998 2008 DH Instruments a Fluke Company 4 4 M MA AI IN NT TE EN NA AN NC CE E A AN ND D A AD DJ JU US ST TM ME EN NT TS S 4 1 MAINTENANCE GB 75 and GB 152 gas booster packages require no stand...

Page 22: ...GB 75 AND GB 152 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...

Page 23: ...position If not set the valve fully to the ON position Check that drive air pressure supply is actually present at the DRIVE ADJUST regulator inlet connection If not ensure drive air is supplied at th...

Page 24: ...ow 2 MPa 300 psi If the supply is too low increase supply pressure Speed of booster pressure generation is directly related to the pressure of the high pressure instrument gas supply For example press...

Page 25: ...at the high pressure circuit connected to the booster OUT port is not open to atmosphere or doesn t have a severe leak If the shop drive air supply pressure is above 500 kPa 75 psig for GB 75 or 250 k...

Page 26: ...eplacement Recommended torque on a DH500 fitting gland is 15 Nm It is possible that a leak exists in the high pressure section of the gas booster These leaks are very difficult to isolate and detect I...

Page 27: ...or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warrant...

Page 28: ...GB 75 AND GB 152 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 22 N NO OT TE ES S...

Reviews: